Viewing a marker pulse – Agilent Technologies E8267D PSG User Manual

Page 122

Advertising
background image

108

Chapter 3

Basic Digital Operation
Using Waveform Markers

The markers are enabled or disabled per your selections, and the changes have been saved to the
selected sequence file.

Viewing a Marker Pulse

When a waveform plays (

page 94

), you can detect a set and enabled marker’s pulse at the rear panel

event connector that corresponds to that marker number. This example demonstrates how to view a
marker pulse generated by a waveform segment that has at least one marker point set (

page 104

).

The process is the same for a waveform sequence.

This example uses the factory–supplied segment, SINE_TEST_WFM in the Dual ARB Player.
Factory–supplied segments have a marker point on the first sample point for all four markers, as
shown.

1.

Press

Mode

>

Dual ARB

>

Select Waveform

, and highlight the desired segment (in this example,

SINE_TEST_WFM

).

2.

Press

ARB Off On

to On.

3.

Connect an oscilloscope input to the EVENT 1 connector, and trigger on the Event 1 signal.

When a marker is present, the oscilloscope displays a marker pulse, as shown in the following
example.

Sequence Marker Column

This entry shows that only
marker 3 is enabled for this
segment.

Marker points on

first sample point of

waveform segment

Viewing markers is described on