Agilent Technologies E8267D PSG User Manual

Page 276

Advertising
background image

262

Chapter 12

Peripheral Devices
N5102A Digital Signal Interface Module

This error is reported when the output FIFO is overflowing in the

digital module. This error can be generated if an external clock

or its reference is not set up properly, or if the internal VCO is

unlocked.

806

Digital module output FIFO underflow error; There are not
enough samples being produced for the current clock rate.
Verify that the digital module clock is set up properly.

This error is reported when the output FIFO is underflowing in the digital
module. This error can be generated if an external clock or its reference is not
set up properly, or if the internal VCO is unlocked.

2.

If the port configuration is parallel or parallel interleaved, using an IQ signal type, press the

Clocks Per Sample

softkey.

Notice that multiple clocks per sample can be selected. Some DACs require the ability to clock
multiple times for each sample; having a clocks per sample value greater than one reduces the
sample rate by a factor equal to the selected number of clocks per sample. The sample rate is
viewed on the first- level and Data Setup softkey menus.

3.

Select the clocks per sample value to fit the test.

4.

Press the

Clock Source

softkey.

From this menu, select the clock signal source. With each selection, the clock routing display in
the signal generator clock setup menu will change to reflect the current clock source. This will be
indicated by a change in the graphic.

5.

Select the clock source.

If External or Device is Selected

Press the

Clock Rate

softkey and enter the clock rate of the externally applied clock signal.

NOTE

The clock phase and clock skew may need to be adjusted each time the clock rate setting is
changed. Refer to

“Clock Timing for Phase and Skew Adjustments” on page 251

.

For the

External

selection, the signal is supplied by an external clock source and applied to the Ext

Clock In connector. For the

Device

selection, the clock signal is supplied through the Device

Interface connector, generally by the device under test.

If Internal is Selected

Using an external frequency reference, the N5102A module generates its own internal clock signal.
The reference frequency signal must be applied to the Freq Ref connector on the digital module.

a.

Press the

Reference Frequency

softkey and enter the frequency of the externally applied frequency

reference.

b.

Press the

Clock Rate

softkey and enter the appropriate clock rate.

Table 12- 8

provides a quick view of the settings and connections associated with each clock

source selection.