56

Chapter 2

Basic Operation
Configuring the RF Output

Figure 2-7

Alternating Sweeps on 8757D

Configuring an Amplitude Sweep

1.

Press

Return

>

Sweep

>

Off

.

This turns off both the current sweep and the alternate sweep from the previous task. The
current CW settings now control the RF output.

2.

Press

Configure Ramp/Step Sweep

.

3.

Using the

Ampl Start

and

Ampl Stop

softkeys, set an amplitude range to be swept.

4.

Press

Return

>

Sweep

>

Ampl

.

The new amplitude ramp sweep settings control the RF output and the CW mode is turned off.

Configuring a Ramp Sweep for a Master/Slave Setup

This procedure shows you how to configure two PSGs and an 8757D to work in a master/slave setup.
The master/slave control setup must use two instruments from the same signal generator family such
as two PSG’s, or two 83640B’s, or two 83751B’s.

NOTE

The master/slave setup applies to ramp sweep only, not step sweep or list sweep. To use this
setup, you must have two sources from the same signal generator family. For example, you
cannot us a PSG with an 83640B or a PSG with an 83751B.