Control Company 4331 ANEMOMETER/THERMOMETER User Manual

Traceable, Anemometer/thermometer instructions, Control company

Advertising
background image

points have been measured and recorded. A

maximum of 8 points may be recorded at one

time.

3. Once all measurements have been recorded,

press the AVERAGE MULTIPOINT button. The

unit will display the average air velocity

reading and number of points measured.

Auto Power Off

The unit will turn off automatically after 15

minutes of non-use to save the battery. This

action is preceded by 3 beeps.

Non-Sleep Mode (Bypass Auto Power Off)

Power the unit off. Press ON and HOLD buttons

at the same time. Release the ON button only. An

“n” indicator appears on display. Release HOLD

button. The unit will remain on and in Bypass

Auto Power Off mode until OFF button is pressed.

OPERATIONAL DIFFICULTIES

1. If the display is flashing or there is no display

when switched on, change the battery (see

Battery Replacement section).

2. If the display indicates an “E6” reading, the

related circuits or parts of thermistor have

failed (see Warranty, Service, or Recalibration

section).

3. If the unit’s fan will not turn this indicates that

the sensor fan is damaged (see Warranty,

Service, or Recalibration section).

4. If the unit does not function properly for any

reason, other than those listed above, replace

the battery with a new high-quality battery (see

Battery Replacement section). Low battery

power can occasionally cause any number of

apparent operational difficulties. Replacing the

battery with a new fresh battery will solve most

difficulties.

BATTERY REPLACEMENT

Erratic readings, a faint display, a flashing display

or no display are all indicators that the battery

must be replaced. To replace the battery, remove

the screw from the lower panel on the back of the

unit. Lift the panel out and remove the exhausted

battery and replace with a new 9-volt alkaline

battery and replace the cover. Replacement

battery Cat. No. 1112.

WARRANTY, SERVICE,

OR RECALIBRATION SECTION

For warranty, service, or recalibration, contact:

TRACEABLE

®

ANEMOMETER/THERMOMETER INSTRUCTIONS

CONTROL COMPANY

4455 Rex Road

Friendswood, TX 77546 USA

Ph. 281-482-1714

Fax 281-482-9448

E-mail [email protected]

www.control3.com

Control Company is

ISO 9001 Quality-Certified by DNV

and ISO 17025 accredited

as a Calibration Laboratory by A2LA.

Cat. No. 4331

Traceable® Anemometer/Thermometer

Instructions

92-4331-00

Rev. 1 071504

Traceable® is a registered trademark

of Control Company

© 2004 Control Company

SPECIFICATIONS

Display: ½

4-digit, 2-line LCD

Range: Air Velocity

Meters - 0.40 to 30.00 per sec.

Kilometers - 1.4 to 108.0 per hour.

Feet - 80 to 5900 per minute.

Miles - 0.9 to 67.0 per hour.

Knots - 0.8 to 58.0

Temperature

14.0° to 122°F and 10.0° to 50.0°C

Resolution:

Air Velocity

Meters - 0.01 meter

Kilometers - 0.1 km

Feet - 1 foot

Miles - 0.1 mile

Knots - 0.1 knot

Temperature - 0.1° (for both °F and °C)

Accuracy:

Anemometer - ±3% of full scale

Thermometer - ±1°C

Size:

7¼ x 3 x 1¾

(fan diameter 2¾

″)

Weight: 7.3 oz

FEATURES

ABS plastic case, chemical & shock resistant

Extra-large digits for easy readings in difficult

locations

Single hand operation

Display updates instantaneously

Memory button permits recalling highest and

lowest readings

HOLD button freezes the display at the current

reading

AVERAGE button allows from 2 to 8 readings

to be averaged automatically

Single point air velocity measurement for

minimum, maximum and average readings

Continuous moving average air velocity

measurement

Auto power off function

Non-sleep mode (bypass auto power off)

Unit selectable (Imperial/Metric): ft/min, mph,

knots / m/s, km/hr, knots

Supplied with RS-232 serial output jack to

connect directly to a computer

RS-232 interface setting

BUTTONS/DISPLAY/INDICATORS

1. ON-OFF/RESET button

2. HOLD button - Freezes the display

3. AVERAGE MULTI-POINT button - computes

average for multi-point readings

4. SEL: MIL/H KNOT button - Selects required

air velocity units

5. MN/MX REC SINGLE-POINT button -

Selects minimum, maximum and average air

velocity for a single point reading.

6. Primary display - Air velocity

7. Secondary display - Temperature display or

indicator for number of multipoint readings in

Multi-Point mode

8. “vel” Indicator - always on for air velocity

measurement

9. “Hold” - Indicates display is frozen

10. “knots” - Nautical miles per hour = 1850

meters per hour = 1.15 miles per hour

11. “ft/m” or “m/s” - Feet per minute or meters

per second

12. “mil/h” or “km/h” - Miles per hour or

kilometers per hour

13. “

°C” - degrees Celsius

14. “

°F” - degrees Fahrenheit

15. “REC”-Indicates reading is recorded and

saved

16. “MIN” - Minimum reading indicator

17. “MAX” - Maximum reading indicator

18. “AVG” - Average reading indicator

19. “

−“ - Minus temperature reading indicator

OPERATION

Measurement Default (Imperial /Metric)

1. Make sure the unit is OFF by pressing the

ON-OFF/RESET button.

2. Press and hold the AVERAGE button. Then press

the ON/OFF button once to turn the unit on. When

the “ft/m, m/s” and “

°

C,

°

F” indicators appear,

release the AVERAGE button.

3. To select metric units, press the HOLD button. “m/s

and “

°

C” will show on the display. To select

imperial units, press the AVERAGE button. “ft/m”

and “

°

F” will appear on display.

4. Press the MIN/MAX REC button. A “5” will show on

the display. Press the HOLD button. A pre-setting

of “1200” or “2400” baud will display.

5. Press the MIN/MAX REC button again. The “5”

indicator will display again. Press the HOLD

button; the unit will revert to the normal measure-

ment display. The default setting is now complete.

6. Press the HOLD button to measure the current air

velocity and temperature with the selected units.

The “HOLD” indicator should disappear.

NOTE: Not pressing the HOLD button retains the

display in freeze mode.

Air Velocity Measurements

ft/min, mph or knots - Set the unit to Imperial

measures. Each press of the “SEL: MIL/H KNOT”

button will change the reading to ft/min, mil/h or knots.

Km/hr or m/s or knots - Set the unit to metric

measures. Each press of the “SEL: MIL/H KNOT”

button will change the reading to Km/hr, m/s or knots.

Typical Air Velocity and Temperature Measurement

Place the meter in an air stream. Make sure that the air

stream and the sensor are aligned. Allow three

seconds for the reading to stabilize.

Measuring Air Velocity (Single Point)

1. Press ON/OFF button to turn on unit. The display

will appear after 5 seconds.

2. Unit is ready for use when the “vel” indicator

appears at upper left hand corner and “

°

C” or “

°

F”

appears at the lower right hand corner.

Continuous Moving Average

The unit can display a continuous moving average for

up to 2 hours.

1. Power the unit ON.

2. Place the unit in an airflow.

3. Press MIN/MAX REC SINGLE-POINT button once.

“REC” and “AVG” appearing in the lower left corner

of the display confirms that the meter is in

continuous moving average measurement mode.

The display will update once per second.

Minimum/Maximum/Average Reading

On a Single Point

1. Power the unit ON.

2. Place the unit in an airflow.

3. Press the MIN/MAX REC SINGLE-POINT button.

The unit will begin to record the readings. The

meter displays the average air velocity by default.

Each press of the MIN/MAX button cycles the

display through the following readings:

No indicator - Real-time readings

REC - Unit is recording the reading

AVG REC - Unit is recording the average velocity

MIN REC - Unit is recording minimum velocity

MAX REC - Unit is recording maximum velocity

4. Press the HOLD button to store the readings before

moving the meter away from an airflow source.

5. Press ON/OFF RESET to turn the meter OFF.

6. To clear the current MIN/MAX average readings,

turn off the power or press and hold MIN/MAX

button until the unit beeps twice, then release.

Air Velocity Average for Multiple Points

1. Turn the meter on and place the unit in an airflow at

the first point to be measured. As soon as the first

measurement is completed, press the HOLD

button, (you will hear a single beep) and release.

The display will show HOLD above the reading.

2. Press the MIN/MAX button, (a single beep will be

heard) and release (the secondary display will show

a digit). This number represents the point that is

being recorded. Repeat this process until all desired