SPECIFICATIONS

Range:

–58.0 to 1832 °F

–50.0 to 1000 °C

Resolution:

0.1° (-9.9 to 199.9) 1° (otherwise)

OPERATION

Press and continue to hold down the trigger to take temperature readings.
The unit will take temperature readings while the trigger is pressed and held
(SCAN appearing on the display indicates measurements are being taken).

After the trigger is released, the last temperature measured will be displayed
for approximately 7 seconds and the display will turn off automatically.
HOLD will appear on the display when the trigger has been released to
indicate that the unit is no longer taking temperature measurements.

SELECTING UNITS OF MEASURE

While taking measurements, press
the °F button to display the temperature in °F, press the °C button to display
the temperature in °C. Once the unit has turned off, the next time the trigger
is pressed, unit of measure will return to the default of °C.

LASER TARGETING

While taking measurement, press the

button to activate laser targeting.

Aim the laser approximately ½ inch above the desired measurement point.

To turn off laser targeting, press the

button again.

CAUTION: Eye damage may result from direct exposure to laser light.

BACKLIGHT

While taking measurements, press the

button to turn on the LCD

backlight. Press the

button again to turn off the LCD backlight.


DISTANCE SPOT/RATIO

There are other factors that may affect measurement accuracy. The target
must completely fill the spot diameter seen by the infrared sensor; otherwise
readings will be influenced by the surface surrounding the target. The ratio of
the distance to the size of the spot being measured is 50:1. For example, an
object’s diameter of 100mm can be measured from 5000mm.

DISPLAY MESSAGES

1” will appear on the display when the temperature being measured is

outside the range of the unit.

ALL OPERATIONAL DIFFICULTIES

If this thermometer does not function properly for any reason, replace the
battery with a new high quality battery (see Battery Replacement section).
Low battery power can occasionally cause any number of “apparent”
operational difficulties. Replacing the battery
with a new fresh battery will solve most difficulties.

BATTERY REPLACEMENT

Erratic readings, faint readings, no display, or a battery symbol appearing on
the display are all indications that the battery must be replaced.




Locate the indentions on either side of the
plastic case near the trigger. Open the
battery compartment by pulling the rubber
portion of the handle out and away from the
trigger. Pull hard, it requires a good amount
of force to open the battery compartment.
Remove the exhausted battery and replace it
with a new 9-volt alkaline battery. Close the
battery cover by snapping it shut.
























WARRANTY, SERVICE, OR RECALIBRATION

For warranty, service, or recalibration, contact:

CONTROL COMPANY

4455 Rex Road

Friendswood, Texas 77546 USA

Ph. 281-482-1714 Fax 281-482-9448

E-mail [email protected]

www.control3.com

Control Company is ISO 9001 Quality-

Certified by DNV and ISO 17025 accredited

as a Calibration Laboratory

by A2LA.

TRACEABLE® INFRARED WIDE-RANGE THERMOMETER

50:1 FIELD OF VIEW INSTRUCTIONS

Advertising