Troubleshooting – Nova-Tech Chem-Tech Series 100D-150D User Manual

Page 10

Advertising
background image

10

TROUBLESHOOTING

PROBLEM

PROBABLE CAUSE

REMEDY

LOSS OF

CHEMICAL

RESIDUAL

1. Pump setting too low.

2. Scale at injection point

3. Solution container allowed to

run dry

1. Adjust to higher setting (feeder must be operating during the stroke length

adjustment).

2. Clean injection parts with 8% muriatic acid or undiluted vinegar.

3. Refill the tank with solution and prime. See Start-Up Section

TOO MUCH

CHEMICAL

1. Pump setting too high.

2. Chemical in solution tank too

rich.

3. Siphoning of chemical into

well or main line

1. Lower pump setting (pump must be operating to adjust the dial).

2. Dilute chemical solution. NOTE: For chemical that reacts with water, it may

be necessary to purchase a more dilute grade of chemical direct from

chemical supplier.

3. Test for suction or vacuum at the injection point. If suction exists, install an

anti-siphon valve. See Figure G.

LEAKAGE

AROUND TUBING

CONNECTIONS

1. Worn tube ends

2. Chemical attack

1. Cut off end of tubing (about 1") and then slip on as before or replace

suction valve housing and compression fitting to prevent leakage.

2. Consult your chemical supplier for compatible materials.

FAILURE TO

PUMP OR FEED

1. Leak in suction side of pump.

2. Valve seats not sealing.

3. Low setting on pump.

4. Low solution level.

5. Diaphragm ruptured.

6. Pump head cracked or

broken.

7. Pump head contains air or

chlorine gas.

1. Examine suction tubing. If worn at the end, cut approximately an inch off

and replace or replace valve body and coupling nut.

2. Clean valve seats if dirty or replace with proper material if deterioration is

noted.

3. When pumping against pressure, the dial should be set above 40%

maximum rated capacity for a reliable feed rate.

4. Solution must be above foot valve.

5. Replace diaphragm as shown in "Service" Section. Check for pressure

above 100 PSI (7 BAR) at the injection point. NOTE: Chemical incompatibility

with diaphragm material can cause diaphragm rupture and leakage around

the pump head.

6. Replace pump head as shown in "Service" Section, Do not use pipe tape

or other sealants. Make sure fittings hand tight only. Using pliers or wrench

can crack pump head. Also, chemical incompatibility can cause cracking and

subsequent leakage.

7. While pump is running, turn bleed valve adjustment screw counter-

clockwise until air is purged. Close bleed valve.

PUMP LOSES

PRIME

1. Dirty check valve.

2. Ball checks not seating or not

sealing properly.

3. Solution container allowed to

run dry

1. Remove and replace or clean off any scale or sediment.

2. Check seat and ball checks for chips, clean gently. If deformity or

deterioration is noted, replace part with proper material. Chemical

crystallization can hold check valves open, therefore the valves must be

disassembled and cleaned. Be sure to replace all parts as shown in the Parts

Diagram (at the end of the manual).

3. Refill the tank with solution and prime.

LEAKAGE AT

FITTING

1. Loose fittings

2. Broken or twisted gasket

3. Chemical attack

1. All fittings can be hand tightened to prevent leakage. Clean off chemicals

which have spilled on pump.

2. Check gaskets and replace if broken or damaged.

3. Consult your chemical supplier for compatible materials.

PUMP WILL NOT

PRIME

1. Too much pressure at

discharge

2. Check valves not sealing

3. Output dial not set at

maximum

1. Open bleed valve and circulate fluid until all air is purged from pump head

assembly. Close bleed valve.

2. Disassemble, loosen, clean and check for deterioration or swelling.

Reassemble and wet the valve assembly, then prime. See INSTALLATION

Section.

3. Always prime pump with output dial set at maximum rated capacity.

ANTI-SIPHON

VALVE

MALFUNCTION

1. Scale or particles have

plugged diaphragm

2. Ruptured valves

1. Remove, clean and reassemble, being careful not to wrinkle the

diaphragm. Check sequence and position of parts to be sure reassembly is

correct.

2. Consult your distributor for replacement.

PUMP MOTOR

STALLS

1. Pumping against excessive

pressure

2. Low voltage to pump

1. Test pressure to determine if it exceeds pump specifications. If so, consult

your distributor.

2. Make sure voltage of power source matches the voltage on the pump

specifications label. If not transformers are available.

MOTOR RUNNING

VERY HOT

1. Low voltage.

2. If using a step-down

transformer, it may be

undersized for the pump

1. Power supply voltage should match voltage on pump specification label.

2. Check the transformer to be sure it has at least 100 watts capacity.

Advertising