PASCO ES-9070 COULOMB BALANCE User Manual

Page 18

Advertising
background image

Coulomb Balance

012-03760E

14

1

. a. Use the allen wrench to

loosen both set screws
one full turn.

b. Thread the wire in the

direction shown.

c. Do not tighten the screws.

8

. a. Using the torsion wire

clamp and a spring
balance, pull the wire
with a force of
approximately 4
newtons (400 gram
mass).

b. Tighten the thumb-

screw.

5

. a. Carefully lay the torsion

balance on its side.

b. Thread the wire through

the hole in the torsion
knob. It is helpful to hold
the thread against the
small allen wrench,
pushing them through
together, as shown.

allen wrench

4

. Thread the wire

through the washer.

3

. a. Remove the wire retainer by

unscrewing it.

b. Loosen the set screw and

thumbscrew one full turn.

c. Thread the wire in the

direction shown.

d. Retighten the set screw, but

do not tighten the thumb-
screw. The wire must be
able to turn.

Torsion wire

9

. Adjust the height of

the pendulum assem-
bly, then tighten the
set screws with the
allen wrench.

2

. a. Using the allen wrench,

loosen the set screw one
full turn.

b. Loosen the thumbscrew

one full turn.

c. Thread the wire in the

direction shown.

d. Tighten both screws.

6

. Put the lower wire retainer

back in position, sliding the
wire through the slot in the
bottom bracket.

7

. Screw the wire

retainer back in
place.

Figure 9. Replacing the Torsion Wire

Advertising