Perma Pure MDH-Series User Manual

Page 2

Advertising
background image

If condensation forms in the unit:
- Decrease or stop sample flow and increase
purge flow until unit dries out

If unit does not heat:
- Check wiring connections
- Confirm power source is operating

(Refer to Figure 1)

1. Connect the 2 wires to a 110 VAC power

source and let the unit heat for 15 minutes

2. Secure sample inlet and outlet connections

using insulated fitting for sample inlet

3. Connect purge inlet and outlet fittings

4. Start purge flow(purge air must be -40°C dew

point air, nitrogen or other gas)

5. Start sample flow. Sample must be maintained

at constant temperature from its origin to
prevent condensation

WARNING

Do not obstruct vent holes

INSTALLATION GUIDELINES

TROUBLESHOOTING

When installing MDH-Series gas dryers, the following
rules apply:

- Sample pressure higher or equal to purge pressure

- Sample gas pressure not to exceed 80 psig and 10

psig for purge gas

- Purge air of -40°C dew point at a flow rate of 15

times sample flow

- Purge air must flow countercurrent to sample

Model Number: MDH-110-96F-4

Nafion Tube OD/ID: 0.108"/0.086"

Length: 96 inches

Nafion Dryer Housing: Fluorocarbon

Enclosure: Polycarbonate

Thermostat Set Point: 80° C

Power Requirements: 110 VAC, 0.3 Amps,
30 Watts

Figure 1

SETUP

HEX NUT

HEX NUT

Model Numbers:
MDH-070-48F-410,

MDH-070-48F-412

MDH-070-72F-410, MDH-070-72F-412
MDH-070-96F-410, MDH-070-96F-412
MDH-110-48F-410, MDH-110-48F-412
MDH-110-72F-410, MDH-110-72F-412
MDH-110-96F-410, MDH-110-96F-412

Nafion Tube OD/ID: 0.108"/0.086"

Length: 96 inches

Nafion Dryer Housing: Fluorocarbon

Enclosure: Polycarbonate

Thermostat Set Point: 80° C

Power Requirements: 30 watts,
110 VAC, 0.3 A,

or 12 VDC, 2.5 Amps.

1. Connect the 2 wires to the appropriate 110
VAC or 12 VDC power source and let the unit
heat for 15 minutes.

rev. 9/2012

48-96 inches depending on dryer