Water-to-gas performance, Gas-to-gas performance – Perma Pure FC-Series User Manual

Page 2

Advertising
background image

FC HUMIDIFIER SPECIFICATIONS

SETUP 2: WATER-TO-GAS

SETUP 1: GAS TO GAS

In Water-to-Gas setup, heated water needs to be
continuously circulated. Flow with greater pressure
needs to be flowing inside tubes to prevent tubing
collapse. DI water should be circulated at 4 LPM for
50 LPM of dry gas flow.

Performance can be controlled by increasing heat
or by decreasing sample flow. Chart below shows
performance of various models.

WATER-TO-GAS PERFORMANCE

Water to Gas Humidification using FC-Style Humidifiers

30

35

40

45

50

55

60

65

70

75

80

200-10"

232

464

696

929

1,161

1,393

1,625

1,857

2,089

200-7"

163

325

488

650

813

975

1,138

1,300

1,463

150-10"

143

286

429

571

714

857

1000

1143

1286

150-7"

100

200

300

400

500

600

700

800

900

125-10"

71

143

214

286

357

429

500

571

643

125-7"

50

100

150

200

250

300

350

400

450

125-5"

36

71

107

143

179

214

250

286

321

Dew Point (C)

Outlet DP (75C water)

Outlet DP (60C water)

Outlet DP (40C water)

Outlet DP (80C water)

GAS-TO-GAS PERFORMANCE

In Gas-to-Gas setup, constant depoint wet gas needs
to be continuously flowing at same flowrate as
sample. DI water is recommended.

B

A

Model Number

4 Deg.

6 Deg.

9 Deg.

FC125-240-5PP

13

30

40

FC125-240-7PP

18

42

56

FC125-240-10PP

25

60

80

FC150-480-7PP

35

84

112

FC150-480-10PP

50

120

160

FC200-780-7PP

57

137

182

FC200-780-10PP

80

195

260

FC300-1660-7ABS

121

291

387

FC300-1660-10ABS

173

415

553

FC300-1660-10HP

173

415

553

FC300-1660-15ABS

259

623

830

FC400-2500-10ABS

260

625

833

FC400-2500-12ABS

312

750

1000

FC400-2500-15ABS

390

938

1250

FC400-2500-20ABS

520

1250

1667

FC600-7000-8PP

583

1400

1860

FC600-7000-14PP

1020

2450

3267

Approach Temperature*

Model Number

A

B

Ports

FC125-240-5PP

8.25"

4.75"

1/2"

FC125-240-7PP

10.25" 6.75"

1/2"

FC125-240-10PP

13.25" 9.75"

1/2"

FC150-480-7PP

11.50" 6.75"

1"

FC150-480-10PP

14.50" 9.75"

1"

FC200-780-7PP

11.25" 6.50"

1"

FC200-780-10PP

14.25" 9.50"

1"

FC300-1660-7ABS

12.10" 5.60"

1.5"

FC300-1660-10ABS 15.10" 8.60"

1.5"

FC300-1660-15ABS 20.10" 13.60"

1.5"

FC400-2500-10PP

16.63" 8.63" 1" or 2"

FC400-2500-12PP

18.63" 10.63" 1" or 2"

FC400-2500-15PP

21.63" 13.63" 1" or 2"

FC400-2500-20PP

26.63" 18.63" 1" or 2"

FC600-7000-8PP

15.5

7.25

2"

FC600-7000-14PP

21.5

13.25

2"

* Units are expressed as flow rate of air (slpm) in
relation to approach dewpoint temperature. Approach
temperature is defined as difference between wet gas
entering humidifier and humidified stream outlet. For
example, for FC-125-240-10PP at 25 lpm, wet gas
enters humidifier at 70 deg.C at Approached Tem-
perature of 4 deg., humidified stream outlet would be
66 deg. C.

Advertising