Micro-GASS User Manual

APPENDIX C CAL 3300 Temperature Controller Configuration

1. For all Micro-GASS systems with the CAL3300 Temperature controller, perform the following setup

procedures.

a. Power Up - Apply power to the system and wait for the controller to perform its self test routine.

The display will then be alternately flashing “nonE” and “inPt”.

b. Input sensor type - Press and hold

Ø key. Press p

key repeatedly to select the “tc J” option for

J type thermocouple. Release both keys.

c. Input unit type - Press the

p

key once. The display will then be alternately flashing “unit” and

nonE”. Press and hold the

Ø key. Press p

key repeatedly to select the “

O

C” unit. Release both

keys.

d. Input setpoint 1 type - Press the

p

key once. The display will then be alternately flashing

SP1.d” and “nonE”. Press and hold the

Ø

key. Press

p

key once to select the output device

SSd”. Release both keys.

e. Input the maximum temperature setpoint - Press and hold the

q

key until “LEVL” appears.

Press and hold the

Ø

key. Press the

q

key to move to level 2. Release the two keys. Press and

hold the

p

key until “hi.SC” appears. Press and hold the

Ø

key. Press the

q

key until “65

appears. Release the two keys.

f.

Input the minimum temperature setpoint - Press the

p

key once. “Lo.SC” appears. Press and

hold the

Ø

key. Press the p

key until “45” appears. Release the two keys.

2. Configure the low temperature alarm.

a. Configure SP2 relay as alarm - Press and hold

p and q

keys simultaneously for 3 seconds to

enter the program mode. Release the two keys. The display will then be alternately flashing
tunE” and “oFF”. Press the

q

key once. The display will then be alternately flashing “LEVL” and

1”. Press and hold the

Ø

key. Press the

p key once to move to level 2. Release the two keys.

Press and hold the p

key until “SP2.A” appears. Press and hold the

Ø

key. Press the

p

key

until the display reads FS.Lo. Release the two keys. The display will then be alternately flashing
SP2.A” and “FS.Lo”.

b. Save configuration into memory - Press and hold

p

and

q

keys simultaneously for 3 seconds.

The display will then be alternately flashing the process temperature and “PArK” indicating that
no setpoint temperature has been entered.