Perma pure llc – Perma Pure Baldwin-Series M8B User Manual

Page 5

Advertising
background image



5

COPYRIGHT 2008, ALL RIGHTS RESERVED

PERMA PURE LLC

Model M8B User’s Manual
Doc. #580: Revision: 000

Page 5 of 22

Figure 2: Heat Exchanger, Impinger and Heat Sink


The Tester’s Choice M8 system will reduce the sample gas dew point to 5

°

C at a maximum flow of 8

Liters Per Minute (LPM). This is slightly above the temperature that the condensate removed from
the gas stream will freeze. If the condensate freezes sample analysis and gas stream flow will be

compromised.

If a lower dew point is necessary, a Perma Pure LLC Supplemental Drying System (S.D.S.) may be
necessary. A sample gas stream that contains significant amounts of SO3 may also be

encountered. Perma Pure LLC/Baldwin can supply SO3 Aerosol Removal Systems, specifically for
this purpose.

Please contact Perma Pure LLC for suggestions on alternative devices. Perma Pure LLC has a
system and expertise that can handle nearly any sample conditioning application.