Teledyne 311D - Portable trace oxygen analyzer with digital meter User Manual

Page 11

Advertising
background image

1. Disconnect

cell.

2.

Move range switch to “CAL” position.

3.

Adjust P1 such that the output of A3, pin 6, measures between 0 and +0.5 mV, ideally

+-.3

mV.

4.

Adjust P2 for 0

± 1 mV at output of A2, pin 6.

5.

Verify that the offset is the same on all ranges.

6. Re-connect

cell.


6. LEAK

TESTING


If a leak is suspected in the unit, DO NOT ATTEMPT TO TIGHTEN THE

DISCONNECT FITTINGS. THE FITTINGS ARE POTTED IN EPOXY AND
TIGHTENING THEM WILL BREAK THE SEAL!
To check for leaks, TAI recommends
one of the following procedures:

Procedure 1:

1.

Purge the instrument down as low as possible.

2.

Place the vent line in water and disconnect the sample.

3.

Next, disconnect the vent line and place the range switch on the X100 range.

4.

The unit should stay on the X100 range if there are no leaks.


Procedure 2:

1.

Purge the instrument with Nitrogen at the sample port.

2.

Note the reading once it has stabilized (at least 24 hours on the 0-10 ppm range) and

increase the flow rate.

3.

If the reading goes down, the unit, or the tubing to the unit, has a leak.



-9-


SPECIFICATION DATA

TAI SALES ORDER NUMBER


INSTRUMENT MODEL NUMBER 311D


INSTRUMENT SERIAL NUMBER


MICRO-FUEL CELL CLASS B-2C


ACCURACY:

±0.2% OF SCALE, OR ±1 PPM, WHICHEVER IS GREATER, AT

A CONSTANT TEMPERATURE;

±5% OF READING, OR ±1 PPM, WHICHEVER

IS GREATER, OVER THE OPERATING TEMPERATURE RANGE.