Teledyne TurbO2 - Portable trace oxygen analyzer User Manual
Page 19
15
Portable Trace Oxygen Analyzer
TURBO
2
Teledyne Analytical Instruments
Specifications
Ranges:
0-100 PPM oxygen
0-1,000 PPM oxygen
0-10,000 PPM oxygen
Sensitivity:
0.5% of full scale
Accuracy:
±2% of full scale at constant temperature;
±5% of reading or ±1 PPM (whichever is
greater) over the operating temperature
range.
Response Time:
90% of full scale in 5-10 seconds at 68°F
(20°C)
Operating Temperature:
+32°F to +125°F (0°C to +52°C)
Micro-fuel Cell Class:
S-2
Micro-fuel Cell Warranty:
6 months
Recommended Span Gas:
Atmospheric air
Operating Power:
Two rechargeable NiCad batteries
Battery Recharge Circuit:
110 VAC, 60 Hz
100/220 VAC, 50/60 Hz optional
Wetted Parts:
Nylon sensor housing and nickel-plated
brass quick-disconnect fittings
Stainless steel optional
Weight:
7 lbs (3 kg)
Dimensions:
10 ¹/
16
" W × 7½ " H × 6¾ " D
(255.6 mm × 190.5 mm × 171.45 mm)
Spare Parts List
Qty
P/N
Description
1
F51
Micro Fuse, 1/2A (220V use F39)
1
C6689-S2 Micro-fuel Cell, Class S-2
2
B76
Battery
1
L79
Lamp
1
A36289
Calibration kit
1
B64442
Pump Assy. (for Pump Option)
A minimum charge is applicable to spare parts orders.
IMPORTANT:
Orders for replacement parts should include the part number
and the model and serial number of the system for which the
parts are intended.