9

1.3.1 Specifications for the QE12, QE 25, QE50, QE65 and QE 95 MB

Footnotes

Model

QE12LP-S-MB
QE12LP-H-MB

Optical Absorber

MB

Spectral Range

1, 8

0.19

– 20 µm

Calibrated Spectral Range
(Optional)
(QE-QED Series)

1, 8

0.248

– 2.5 µm

10.6 µm

0.3

– 2.1 µm

Typical Sensitivity

2, 7

60 V/J

Calibration Uncertainty

7, 2, 8, 9, 11

± 3%

Repeatability

< 0.5 %

Max. Pulse Energy 1.064 µm

0.266 µm

With QED @ 1.064 µm

QED @ 0.266 µm

3, 12

0.85 J

0.7 J
3.9 J

0.81 J

Noise Equivalent Energy (NEE) (Typ)

2, 11

0.7 µJ

Max. Repetition Rate

2, 5, 11

300 Hz

Typical Rise Time
(0-100%)

2, 11

550 µsec

Max. Pulse Width (Typ)

2, 6, 10, 11

400 µsec

Max. Energy Density

12

600 mJ/cm

2

@ 1.064µm, 7ns, 10 Hz

500 mJ/cm

2

@ 266nm, 7ns, 10 Hz

Max. Energy Density with QED

16 J/cm

2

@1064nm, 7nsec, Single shot

8 J/cm

2

@1064nm, 7nsec, 10Hz

6 J/cm

2

@532nm, 7nsec, 10Hz

1 J/cm

2

@266nm, 7nsec, 10Hz

Max. Average Power

Detector Alone (QE12xP-S-MB):

With optional Heatsink (QE12xP-H-MB):

12, 13

3 W (7.5 W with QED)

5 W (12.5 W with QED)

Max. Power Density

Detector Alone (QE12xP-S-MB):

With optional Heatsink (QE12xP-H-MB):
With QED :

10 W/cm² @ 3 W
10 W/cm² @ 5 W

600 W/cm²

Dimensions (H x W x D)

Detector Alone (QE12xP-S-MB):

With optional Heatsink (QE12xP-H-MB):

36 x 36 x 14 mm
36 x 36 x 33 mm

Weight : Detector Alone (QE12xP-S-MB):

With Optional Heatsink (QE12xP-H-MB):

87 g

117 g

Aperture Size QE
12:

QED 12:

12 x 12 mm

9 x 9 mm

Aperture Area Size QE
12:

QED 12:

1.4 cm

2

0.81 cm

2