Gentec-EO S-LINK User Manual
Page 30
S-
LINK User’s Manual Revision 2.4
25
2:100.231 : 100.231 Watts on channel 2.
Energy Mode: 4 digits in scientific notation.
2:8.808e-07 : 0.8808 uJ on channel 2
If you have a repetition rate higher than the ASCII specification, the monitor sends you
an error message. If it does, try sending the *SS100x command to ignore some laser
pulses.
By default the monitor is set in binary mode. To set the ASCII mode, send the serial
commands *SS1001 for channel 1 or *SS2001 for channel 2.
*CS1
/
*CS2
To end/exit the *CA1 or *CA2 mode.
*AN1x
/
*AN2x
For Power heads only. To set the Anticipation for Channel 1 or 2.
x= 0 = disable, x = 1= enable
Default value is 1.
Example : Enter *AT10 to disable the anticipation for Channel 1.
Example : Enter *AT21 to enable the anticipation for Channel 2.
*RST
To reset the device. The monitor will read
the head’s EEprom again.
*PW1
/
*PW2
To set the Wavelength for Channel 1 or 2.
Wavelength correction command (5+ characters).
Example : To select the wavelength 1064 nm, enter *PW101064. Example : To
select the wavelength 10600 nm, enter*PW110600.
*TL10002
/
*TL20002
To modify the trigger level in energy mode
*TL1XXXX / *TL2XXXX where XXXX are four characters representing the triggel level
(in percent) needed.
for example :
XXXX = 0002 for a trigger level of 2%
XXXX = 00.5 for a trigger level of 0.5 %
XXXX = 11.4 for a trigger level of 11.4%
XXXX = 0050 for a trigger level of 50%
Note : the "." is indeed a character
*VER
To display the version of the software.