INSTALLATION

INSTALLATION

Figure 4: Installation

MOUNTING

Using four (4) each 10" X 32" or M-5 screws (not

included), secure the front panel of the High Volt-

age Controller, using the supplied mounting holes,

to enclosure.

Figure 5: Input Power Connections

INPUT POWER

CONNECTIONS

> The Ransburg High Voltage Controller

has a built in resettable fuse so if it draws

a current in excess of 2.5 amps it will open.

Reset is achieved by turning controller

power OFF for 5 minutes then back ON.

NOTE

Input power must be supplied from a regulated

DC power Supply. The connection is made at J8

with +VIN at 24VDC ± .5V, GND at 24VDC return,

CGND at Earth Ground. The 24 VDC supply must

be protected against excessive current as well as

Over Voltage protected.

High Voltage Controller - Installation

10

Ransburg

LN-9623-00.1