Troubleshooting – Robinair TIF8800X Combustible Gas Detector User Manual

Page 14

Advertising
background image

12

Troubleshooting

Symptom

Possible Cause

Solution

Power indicator LED

does not light; tool is

non-responsive .

1) Battery pack not charged .

2) AC adapter is connected

to tool .

1) Connect AC adapter to tool

and charge battery pack . Refer

to Setup for instructions .

2) Remove AC adapter .

Fault LED lights during

charging .

1) Problem with battery pack .

2) Internal failure .

1) Disconnect AC adapter from

tool and wait one hour for

battery pack to cool . Do not

use tool or attempt to charge

battery pack during this time .

After one hour, reconnect AC

adapter to tool. If red LED

(fault) illuminates again,

replace battery pack .

Refer to Replacement Parts

section of this manual .

2) If battery pack is not the

problem, discontinue using tool

and return it for service .

CauTiOn: This tool should

be serviced by TiF only. Failure

to do so may impair the

intrinsic safety of the device.

Tool does not seem

operable; does not alarm

in the presence of leaks .

1) Sensor failed .

2) Internal failure .

1) Verify unit on known

combustible gas leak source . If

no response, replace sensor .

Refer to Maintenance for

instructions .

2) If replacing sensor does not

solve problem, immediately

discontinue using tool and

return it for service .

CauTiOn: This tool should

be serviced by TiF only. Failure

to do so may impair the

intrinsic safety of the device.