Terminal arrangement, Standard input ranges – Ronan X54-3402 User Manual

Page 2

Advertising
background image

Page 2 of 2

Rev. 0

Ronan Engineering Company

Ph: 800-327-6626 Fax: 818-992-6435 E-mail: [email protected] Web: www.ronan.com

TERMINAL ARRANGEMENT:

X54-3402

Dual Channel Universal RTD Alarm Trip (continued)

MECHANICAL DETAIL:

Above ranges are standard offerings for

fastest delivery. Contact Ronan for price

and delivery of non-standard ranges.

These standard ranges can be re-

configured in the field or pre-configured at
the factory to the customer’s specific RTD

type and temperature range.

ORDERING INFORMATION:


X54-3402-(I) (II) - (I) (II)

Channel #2
Channel #1

I = RTD type (100 = 100 ohm Platinum 0.00385)
(120 = 120 ohm Nickel)
( 10 = 10 ohm Copper @ 0°C)
( 9 = 10 ohm Copper @ 25°C)
II = Standard Range code (A - G) or
specify special range; e.g. 0-450°F = (0/450°F)

Example: Standard Ranges 0-500°F
X54-3402-100E-100E

Example: Special Range 0-450°F
X54-3402-100(0/450°F) -100(0/450°F)

Accessories:
D2-35X7.5 35 mm X 7.5 mm “U” style DIN rail, sold per foot

*
1.) Plug-in connector design simplifies installation and

maintenance.

2.) Use up to 12 awg wiring.
3.) When inserted, case back cover provides a mechanical

cover for connector screws (as required by European
standards).

Code

ºF

ºC

Sensor

Types

A

-40 to +40

-40 to +4

100, 120

B

-40 to +104

-40 to +40

100, 120

C

32 to +100

0 to +38

100, 120,

10, 9

D

32 to +212

0 to +100

100, 120,

10, 9

E

0 to +500

-18 to +260

100, 120

F

32 to +932

0 to +500

100

G

0 to +1500

-18 to +815

100

STANDARD INPUT RANGES:

0.850”
21.6mm

2.47”
62.7mm

2.23”
54.6mm

5.80”
147.3mm

C

L

DIN Rail

1

2

3

4

5

6

1

2

3

4

5

6

1

2

3

J2*

J4*

J5*

J2*

Outputs

18-30 vdc
Power
Input

+

-

GND

NC
NO

C

NC
NO

C

NC
NO

C

NC
NO

C

B Setpt.

A Setpt.

B Setpt.

A Setpt.

Ch1

Ch2

1

2

3

4

5

6

J1*

Ch1

Ch2

3-wire RTD

Inputs

Advertising