Calibration – SKC 100-3000 Leland Legacy User Manual

Page 18

Advertising
background image

15

6. CCAL will display during Calibration Check mode and will count down

to 1. The pump will stop running.

7. When calibration is completed, the Leland Legacy

will go to Hold. If the calibration was successful, the
pump LCD will revert to displaying pump run time as
0.0. If there was failure during the calibration process,
an error code of E4[x] will appear (see CalChek Error
Chart on page 16
).

Note: To remove a CalChek error code from the LCD, press .

8. Allow the pump to go to Sleep mode to write calibration data to pump

memory.

CalChek Full Calibration Data

Requires DataTrac for Leland Legacy Software
Full calibration completely clears pump history, run time parameters, and the
DataTrac Scheduler. Full calibration data can be viewed and printed by going to
the DataTrac Pump Manager window in DataTrac for Leland Legacy Soft ware
and clicking on the View menu. Choose Calibration Info. This will display
calibration results, pump serial number, and date of the last full calibration. A
butt on allows this data to be printed. The printed report contains pump version,
date printed, and a validation code to perform data verifi cation.

CalChek Full Calibration Data Verifi cation

Requires DataTrac for Leland Legacy Software
To ensure that printed calibration data has not been tampered with, pull
down the Tools menu in the Calibration Info window and choose Confi rm
Validation Code. Enter the data from the printed report, including the
validation code. DataTrac Soft ware will indicate whether the information is
completely valid or if a parameter has been changed.

When entering data to confi rm the validation number, enter the
date in the following format: mmm, dd, yyyy (e.g., Aug 18 2009).

Calibration