Setup, User setup functions – SKC 210-5000 AirChek XR5000 User Manual

Page 11

Advertising
background image

8

User Setup Functions

Function Overview

User setup functions are listed below in the order in which they display. Note that the
CLr function for clearing accumulated run time is only available when accumulated
run time exists.

No Accumulated Run Time

Accumulated Run Time
CLr

ADJ Flow

ADJ Flow

Set Timed Run

Set Timed Run

Set Run Delay

Set Run Delay

Clearing Accumulated Run Time

1. With the pump in a non-running state (no fl ashing blue

LED), press .

2. Press [] at CLr display to clear accumulated run

time. Pump is ready to run.

CLr will not cancel Timed Run or Run Delay time settings (see Canceling a
Timed Run and/or Run Delay).

Changing the timed run and/or run delay settings in user setup functions will

automatically clear accumulated run time.

Changing the fl ow rate in user setup functions will not clear accumulated

run time.

Setting Flow Rate

1. With pump in a non-running state (no fl ashing blue

LED), press .

2. Connect pump inlet to a calibrator.
3. Press  until ADJ and Flow fl ash on display.
4. Press

 to increase or  to decrease fl ow. Dashed

lines will move up and down on the LCD to indicate
direction of adjustment. Flow rate will not display
on the pump LCD. Observe the calibrator for fl ow
reading.

5. Press

[] to accept fl ow setting and to exit user setup functions.

See Calibration for instructions on calibrating pump fl ow rate.

Setup

Note

Note