Operation

Procedure for PowerFlex Activation
1. Plug the power cable into the power jack on the back of the PowerFlex. Plug

the wall cube into a power outlet (100-240 V).

If the LED lights red or green upon plugging the charger into the
wall outlet (no pump att ached), discontinue use of the charger and
return it to SKC for repair.

Procedure for Charging AirChek 2000 Pumps
Att ention AirChek 2000 Users: To prevent batt ery damage, follow the steps below
in the order indicated.
1. Ensure the sample pump is off .
2. Open the protective cover on the pump charging port by turning the cover 90

degrees counter-clockwise.

3. Connect the charging plug end of the charging cable

to the charging jack on the AirChek 2000 with the
keyed edge facing down. Caution: Forcing the plug
into the pack with the keyed edge facing up will cause
damage to the batt ery pack.

4. Insert RJ-45 connector on the charging cable into the

PowerFlex charging port.

5. Allow up to 10 minutes for the LED indicator to light

red. If it does not light, see Troubleshooting, Step 1.

6. The batt ery is fully charged when the LED indicator lights green.
7. Disconnect the RJ-45 connector on the charging cable from the PowerFlex

charging port.

8. Disconnect the charging plug end of the cable from the charging jack on the

AirChek 2000 batt ery.

9. Close the protective cover on the pump charging port.

Procedure for Charging Other SKC Pump Models
1. Ensure the sample pump is off .
2. Select the appropriate pump-specifi c cable.
3. Insert the RJ-45 connector into the charging port on the PowerFlex. The connec-

tor should slide easily into the port.

4. Insert the charging plug end of the cable into the charging jack on the appro-

priate sample pump.

AirChek 52 Note: Plug the charging plug end of the cable into the jack on the

back of the batt ery pack.

222 Pump Note: Push the charging plug end of the cable into the charging

jack on the pump until it clicks. This may require some
force.

2

• PowerFlex line source is UL Listed for electrical safety, but the low-voltage

charger unit is not. Do not use PowerFlex in hazardous locations, in rain, or
in standing water. For indoor use only.

Remove pump-specifi c cable from the charger when not in use to avoid

shorting.

Caution

3

Advertising