SKC IOM Calibration Adapter User Manual

Operating instructions

Advertising
background image

Calidaptor

Cat. No. 391-01

To use the 391-01 Calidaptor with an IOM Sampler:

1. Push IOM through hinged bracket and place inlet against foam ring.

2. Clamp IOM in place with plastic clamping screw until foam ring

compresses by 1 mm. Ensure IOM inlet is centered.

3. Screw supplied hose barb into threaded hole in calibration adapter

inlet.

4. Use a length of fl exible tubing to connect hose barb to calibrator outlet.

5. Connect IOM outlet to the inlet of a sample pump.

6. Adjust

pump

fl ow until desired fl ow reading appears on fl owmeter.

For additional information, contact SKC Technical Support

at [email protected]

SKC is an ISO 9001 Registered Manufacturing Company

SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete
SKC Limited Warranty and Return Policy, go to http://www.skcinc.com/warranty.asp.