SKC 225-9820 VersaTrap Sampling Cassette User Manual

Versatrap spore trap cassettes, Operating instructions

Advertising
background image

VersaTrap Spore Trap Cassettes

Cat. Nos. 225-9820 and 225-9821

VersaTrap Spore Trap Cassett es

1. Select

a

fl ow rate (see Flow Rate Guide on back) and calibrate pump fl ow rate.

VersaTrap

®

Cassett es do not produce signifi cant back pressure, therefore, it is

not necessary to calibrate fl ow with a cassett e in line.

2. Remove and retain sealing tabs from cassett e inlet and outlet.

3. Connect

cassett e outlet to pump inlet.

a. Direct method: If pump accommodates direct mounting, insert cassett e

adapter into pump inlet and push cassett e outlet onto bowl end of
adapter.

b. Indirect method: Use 1/2-in ID tubing to connect cassett e outlet to pump

inlet.

4. Turn on sample pump and sample for appropriate sampling time. See

Sampling Interval Guide on back.

5. When sampling is complete, turn off pump, remove cassett e, and reseal with

sealing tabs.

6. Place sample cassett es, outdoor background samples, non-complaint

area samples, and blanks in a corrugated box and ship to a laboratory for
analysis.

Flow Rate Guide

Select the fl ow rate that will provide the desired 50% cut-point to capture the
particle size of interest.

15 L/min is recommended as an optimal fl ow rate.

Sampling Interval Guide

Following are recommended sampling times for specifi c dust conditions.

SKC recommends reducing length of sampling time when using a
30 L/min fl ow rate to prevent overloading.

Form #40063 Rev 1102

Flow Rate

50% Cut-point (μm)

VersaTrap

30.0

1.5

25.0

1.7

20.0

1.9

15.0

2.3

10.0

2.8

5.0

3.9

Recommended Sampling Intervals for the VersaTrap Air Sampling Cassette

at Typical Collection Flow Rates

Typical Environmental
Conditions

Flow rate

Collection

@ 15 L/min

Flow rate

Collection

@ 30 L/min

Clean “offi ce” or
outdoors (no visible dust)

10

minutes

5

minutes

Indoor environment,
high activity personnel

5

minutes

2

minutes

Indoor environment,
drywall renovation, or
heavy industrial dust


1

minute

1/2

minute

Operating Instructions

863 Valley View Road, Eighty Four, PA 15330 USA

Tel: 724-941-9701 Fax: 724-941-1369 e-mail: [email protected]

SKC Limited Warranty and Return Policy

SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC’s sole liability and the buyer’s exclusive remedy. To view the com-
plete SKC Limited Warranty and Return Policy, go to htt p://www.skcinc.com/
warranty.asp.

Advertising