SKC 770-140 Calibration Span Reference for HAZ-DUST IV and VDM-7500 User Manual

SKC Equipment

Advertising
background image

Calibration Span Reference for

HAZ-DUST IV and VDM-7500

Catalog No. 770-140

Operating Instructions

863 Valley View Road, Eighty Four, PA 15330 USA

Tel: 724-941-9701 e-mail: [email protected]

The Calibration Span Reference light-scatt ering device provides a constant value
(known as a "K" factor). Use the Calibration Span Reference to check factory calibration
of the HAZ-DUST IV and VDM-7500 instruments. Calibration should be checked
the fi rst time the instrument is used, once each month with normal usage, and aft er
an instrument has been dropped or damaged. Instruments should be sent to SKC
annually for recalibration.

Calibration Procedure

1.

Ensure sensor optics are clean and batt ery is fully charged (see instrument operating
instructions
). Calibration must be performed in a clean environment.

2. Activate the instrument LCD by pressing ON/OFF.

3.

Insert the thoracic sampling inlet into the sensor head. Att ach the fi lter cassett e
to the sensor head. Att ach the air intake tubing to the fi lter cassett e.

4. Activate the internal pump as follows:

a. Select Run from Main Menu.
b. Select Run - Continue.
c. Select Yes and allow instrument to run for two minutes to stabilize.
d. Press

Enter aft er at least two minutes have elapsed to stop the pump.

5. Perform the auto-zero process as follows for the HAZ-DUST IV or the VDM-

7500:

a. Insert the zeroing fi lter into the thoracic sampling inlet.
b. Select Auto-Zero from Main Menu. The Auto-Zero screen will appear

briefl y.

c. Select Auto-Zero. The Auto-Zero screen will remain during the process. The

Main Menu will display once the process is complete.

6. Remove the thoracic sampling inlet from the sensor head. Disconnect the air

intake tubing.

7. Insert the Calibration Span Reference into the sensor head (Figure 1 on reverse

side).

www.skcinc.com

Form #40084 Rev 1102

Figure 1. Insert Calibration Span Reference into sensor head.

8. Activate the internal sampling pump by repeating Step 4, a. through d. The

internal sampling pump should remain running.

9. Observe the concentration value shown on the instrument LCD. Calculate

the "K" value using the procedure below. The values should match within
± 10%.

a. Agree within ± 10%, the instrument has passed the calibration check.
b. Do not agree within ± 10%, repeat the above procedure to rule out

error, then contact SKC technical support and return the instrument
for recalibration.

10. Remove the Calibration Span Reference, reatt ach the air intake tubing, and

replace the sampling inlet. Run unit for 2 minutes with zero fi lter and perform
auto-zero.

Calculating the "K" Value

1. Follow Steps 1 through 6 for calibrating the instrument (see reverse side).

2.

Repeat Steps 7 through 9 three times and average the results displayed on the
LCD. That average value is the "K" value. Record the "K" value on the Calibration
Span Reference and use it for subsequent calibrations. Write the serial number
of the instrument on the Calibration Span Reference.

SKC Limited Warranty and Return Policy

SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC’s sole liability and the buyer’s exclusive remedy. To view the
complete SKC Limited Warranty and Return Policy, go to htt p://www.skcinc.
com/warranty.asp.

Advertising