SKC 500-200 UMEx 200 Passive Sampler for Sulfur Dioxide User Manual

Operating instructions, Performance profi le

Advertising
background image

Operating Instructions

863 Valley View Road, Eighty Four, PA 15330 USA

Tel: 724-941-9701 Fax: 724-941-1369 e-mail: [email protected]

Passive Sampler for Sulfur Dioxide

Cat. No. 500-200

Performance Profi le

Sampling Rate:

15.2 ml/min with a relative standard deviation of 16.5%

Validated Concentration
Range:

0.4 to 8 ppm, 15 minutes to 24 hours

Analysis Method:

Solvent extraction and IC analysis with conductivity detection

Lower Detection Limits:

15 minutes: 170 ppb

8 hours: 5.3 ppb
24 hours: 1.8 ppb

Collection Method:

Tape treated with triethanolamine (TEA); built-in blank
included


Shelf-life:

18 months from date of manufacture at ambient temperature

Storage:

Before use: Ambient temperature

After use: Analyze within 3 weeks. Can be stored at ambient
temperature or at ≤ 39.2 F (4 C)

Accuracy: ±

30%

Temperature Effects:

No effect on sampling rate from 20 to 25 C

Humidity Effects:

No effect from 20 to 80% relative humidity (RH)

Wind Velocity Effects:

No effect from 5 to 100 cm/sec

Interferences: None

identifi ed

Dimensions:

3.4 x 1.1 x 0.35 in (8.6 x 2.8 x 0.89 cm)

Weight:

0.38 oz (10.9 gm)

Slide Cover: Yellow

Note: NO