SKC 575 Series Passive Samplers

for Organic Vapors

SKC 575 Series Passive Samplers are small ready-to-use samplers that provide
superior collection of organic vapors without the use of a sample pump. SKC
has scientifi cally validated the sampling rates for many chemicals and other critical
sampling parameters of the 575 Series Passive Samplers to meet OSHA/ASTM/
ANSI requirements. SKC 575 Series Passive Samplers have been identifi ed as a viable
alternative to active sampling in six OSHA methods.

Performance Profi le

Sorbent/Amount:

575-001: Charcoal, 350 mg

575-002: Anasorb

®

747, 500 mg

575-003: Anasorb 727, 300 mg

Housing material: Nylon
Diameter:

1.4 in (3.5 cm)

Length
(including clip):

2.5 in (6.3 cm)

Depth:

0.6 in (1.5 cm)

Concentration range: Varies - dependent upon chemical of interest
Analysis:

Solvent desorption, Gas Chromatography/Flame Ionization
Detector (GC/FID)

Shelf-life:

2 years at < 39.2 F (4 C)

Storage:

Before use: Store at ambient temperature. For low-level
measurements, SKC recommends storage at < 39.2 F (4 C).

Aft er use: For sample storage information, refer to the
method for the chemical of interest. Expedited shipping is
recommended. See Storage.

Sample time:

Validated for 15-min and 8-hr occupational exposure sampling.
Lab and fi eld studies conducted with several compounds show
suitability for 24-hr air sampling.

Sampling rate:

Dependent upon chemical of interest. For compound-specifi c
sampling rates, visit www.skcinc.com and click on Sampling Guides.