Operation – SKC 231, 237, 248 Series Dual Stainless Steel Fitted Bag User Manual

Page 2

Advertising
background image

®

000277

Operation

1. Flush the bag at least 3 times with purifi ed air or nitrogen before use.

2. To

fi ll the bag, connect PTFE tubing from the exhaust port of an air sample

pump to the hose connection on the bag (stem protruding from the top of
the inlet fi tt ing).

3. Ensure that the valve is locked open by tightening the

knurled thumbscrew clockwise until snug. Turn on the
sample pump to fi ll the bag (see Proper Bag Infl ation).

Do not sample compounds that exceed the 225 F (107.2 C)
temperature rating of the fi tt ing O-rings.

When using 237 Series SamplePro FlexFilm bags, do not
sample compounds that exceed 140 F (60 C) temperature.

Do not use sample bags to sample compounds with
boiling points > 249.8 F (121 C).

4. When sampling is complete, turn off the pump and

loosen the knurled thumbscrew a maximum of
2 revolutions.

5. Close the shut-off valve by pulling up on the valve stem.

(Note: Ensure the PTFE tubing is not pulled off while
pulling up on the valve stem.) Tighten the knurled screw
until snug to lock the valve closed.

6. To withdraw samples using a needle and syringe,

carefully insert the needle into the port in the septum
fi tt ing and pierce the septum.

Do not allow the syringe needle to puncture the bag
material when piercing the septum.

2

Advertising