Interferences, Possible causes for invalid tests – SKC 560-002 MethChek Immunoassay Wipe Kit User Manual

Page 8

Advertising
background image

8

Interferences

The MethChek Step 3 Cartridge is designed to react positively to
methamphetamine above the particular kit’s lower limit of identifi cation.

Possible Positive Interferences

Negative Interferences

There are no known negative interferences.

Possible Causes for Invalid Tests

Invalid test results are a rare occurrence. Possible causes for an invalid result can
include: damaged test cartridge, storage at high temperatures, expired test cartridges,
interferences in the sample (strong organics, bleaching agents, extremely high or
low pH), or improper test procedure (cartridge not held fl at during development,
insuffi

cient sample volume). If an invalid test result appears, perform a new test

with a new cartridge.

If no lines appear on the Step 3 Cartridge, it may be due to one of the following:
insuffi

cient volume of extract liquid was added to the Step 3 Cartridge or extract

liquid was added to the sample window instead of the sample well (marked “S”)
on the Step 3 Cartridge.

Compound

Approx. ng Equivalent to

50 ng/ml d-methamphetamine % Cross-reactivity

MDMA (Ecstasy)

< 50

+100

d-methamphetamine

50

100

street methamphetamine
(methamphetamine hydrochloride, unknown purity)

50

100

d,l-methamphetamine

80

62

l-methamphetamine

650

8

d-amphetamine

2500

2

phenylephrine (component of nasal sprays)

2500

2

d-ephedrine

4000

0.3

l-ephedrine

> 5000

< 1.0

d,l-ephedrine

> 5000

< 1.0

d-phenylpropanolamine

> 5000

< 1.0

d,l-phenylpropanolamine

> 5000

< 1.0

d-pseudoephedrine

> 5000

< 1.0

l-pseudoephedrine

> 5000

< 1.0

heroin

> 1,000,000

< 0.005

powder cocaine

> 1,000,000

< 0.005

street cocaine

> 1,000,000

< 0.005

Advertising