Thank you for purchasing a SleepSense sensor

Applying the Sensor

Positioning the Sensor

1. Bend the sensor body into a soft curve, to fit onto the patient’s

upper lip. Position the sensor’s black beads under the nostrils and

away from the face.

2. Align the tip of the oral prong with the center of the mouth. Bend

the prong to the desired position.

3. Draw the wires on either side of the sensor over the patient's ears

and use the clasp under the chin to secure the wires in place.

Recommendations:

•

Tape the wire on both cheeks to reduce strain

and enhance reliability.

•

Check that clear, strong signals are being

transmitted before leaving the patient.

The respiration flow sensor is placed on the patient’s upper lip, under the nostrils.

The sensors are delicate and must be handled with care.

Part type:
Cable:
Recommended filter settings:
Intended for system:
Operating conditions:
Storage temperature:
Operating and storage humidity:

•

SleepSense

®

sensors are for professional use only.

•

SleepSense

®

sensors are for sale by or on the order of a physician only.

•

SleepSense

®

sensors may be used only in conjunction with an approved recording system.

•

SleepSense

®

sensors are only intended for use on healthy skin.

•

Discontinue use if sensor shows signs of wear, damage, or exposed metal.

Cleaning the Sensor

•

Wipe the sensor and cable with a non-corrosive (to plastic) cleanser to clean before use.

•

Make sure the complete sensor assembly is thoroughly dry before reusing it.

•

To sterilize the sensor, follow standard gas sterilization protocols.

•

Do not autoclave or soak the sensor system in disinfectants.

1459 / 1460 (3 thermistors)
200 cm long white PVC insulation lead wire
As recommended in the system’s instructions
Respironics

®

Alice 3 and Alice 4

5ºC (40ºF) -- 40ºC (104ºF)
-20ºC (-4ºF) -- 60ºC (140ºF)
5% -- 95% (Non-condensing)

Indications for use

SleepSense

®

sleep-lab sensors provide a qualitative measure of sounds, respiratory-effort, flow, body

position or limb movements, for recording on an approved data acquisition system. They are intended

for use on children and adult patients who are screened during sleep disorder studies at a sleep

laboratory or the patient’s home.

Technical Specifications:

Description

Comprising of three thermistors, (heat sensitive resistors), the respiration flow sensor is placed under

the patient's nose and produces a signal that is directly proportional to the temperature changes of air

inhaled and exhaled during respiration. The sensor generates a small analog electrical voltage that

provides a clear, reliable indication of respiration airflow. The sensor should be connected to the

appropriate channel on the recorder.

Warnings and Precautions

Limited Warranty

SLP warrants the respiration-flow sensor to be free of defects in materials and workmanship for a period of six months from the date purchased.

The sole liability of SLP and our distributor(s) is limited to replacement or repair of the product at the option of SLP with no charge for parts or

labor if any part is proven to be defective in workmanship, performance, or materials during the warranty period. Under no circumstances shall

SLP or our distributor(s) be liable for any loss of revenues or damage, direct, consequential, or incidental, including loss of profit, property

damage, or personal injury arising from the use of, or the inability to use this product. This warranty is intended only for the original buyer and

is in lieu of all other warranties or previous agreements, expressed or implied. This warranty is rendered void if the product is used for other

than its intended purpose or is subject to abuse, misuse, tampering, neglect,or unauthorized modifications. Use of this product constitutes

acceptance of this warranty in total.

For a complete list of technical specifications please contact us or visit our website.