The cursors dialog – Teledyne LeCroy WaveMaster 8 Zi_Zi-A User Manual

Page 98

Advertising
background image

WaveMaster 8 Zi/Zi-A Oscilloscopes

90

922133-00 Rev A

PLEASE NOTE THE FOLLOWING:

The bottom cursor knob adjusts relative cursors. It does not work
with absolute cursor types.

Previously setup cursor positions are preserved for subsequent
displays. Changing cursor positions or measurement modes are
always available using touch screen or front panel control methods
as previously mentioned.

The Cursors Dialog

Whether using touch screen or front panel control methods (as previously
mentioned) to access cursors, the Cursors dialog is then shown.

C

URSORS

O

N

Mark or unmark the Cursors On checkbox to turn cursors on and off as
desired.

C

URSOR

T

YPE

C

ONTROLS

Select your Cursor Type from Horizontal Abs, Horizontal Rel, Vertical Abs,
and Vertical Rel options shown. Subsequent controls on the dialog are
shown, hidden, enabled, or disabled based on the Cursor Type selection.

Horizontal Abs - Provides a single, dashed, vertical line and cross-
hair marking for the specific cursor location. None of the Show
controls are provided; however, the Position 1 control is enabled on
the right of the dialog.

Horizontal Rel - Provides two, dashed, vertical lines and cross-hair
markings for the specific cursor locations. Since there are two
cursors involved, controls for Show along with Track, and Position 1
and 2 are available on the dialog.

Vertical Abs - Provides a single, dashed, horizontal line and cross-
hair marking for the specific cursor location. None of the Show
controls are provided; however, the Position 1 control is enabled on
the right of the dialog.

Advertising