Horizontal TRDL

Command/Query

DESCRIPTION The

TRDL command defines the horizontal position of the

trigger point in time. With the screen center defined
as zero, the time duration from this zero point to the
trigger point is set. The setting range is determined
by the timebase (TDIV). The TRDL? query returns the set
horizontal position of the trigger point.

COMMAND SYNTAX

TRDL <value>
<value>:= Trigger point setup parameter. The <DECIMAL
NUMERIC PROGRAM DATA> format is used, and the suffix
is also valid.

The range of <value>:
<value> := -500s to +5div (Time/div := 50s/div to
10ms/div)
<value> := -5s to +5div (Time/div := 5ms/div to
10us/div)
<value> := -5ms to +5div (Time/div := 5us/div to
500ps/div)
<value> := -500s to +750s (Trigger mode is STOP)
See Remarks for detail.

QUERY SYNTAX

TRDL?

RESPONSE FORMAT

<value>
<value>:= The <NR3 NUMERIC RESPONSE DATA> format is
used. The currently set trigger point is input here.

REMARKS

TRDL reports command-execution error (CME) when
<value> is not a <DECIMAL NUMERIC PROGRAM DATA>.
For more detail information see below.

1. Setting Range

The setting range and rounding method for the TRDL command are defined as
follows, depending on the waveform acquisition and sweep time.

Trigger Mode

Sweep time

Setting range <value>

50s/div to 10ms/div

-500s to
+5div

*

5ms/div to 10us/div

-5s to +5div

*

SINGLE, NORMAL or AUTO

5us/div to 500ps/div

-5ms to +5div

*

STOP

-

-500s to +750s

*

• The above division value in the setting range is calculated from the following
equation.

(Setting range upper limit) = +5div × (Sweep time)
• Any value exceeding the above range is rounded to the maximum or minimum
permissible value, whichever is nearest.

2. Set resolution

The resolution is changed according to the set sweep time. The resolution is

calculated from the following equation.