Set Up Remote Control

QualiPHY software can be executed from a remote host computer, controlling the oscilloscope
through a LAN Connection. To set up remote control:

• The oscilloscope must be connected to a LAN and assigned an IP address (fixed or

dynamic).

• The host computer must be on the same LAN as the oscilloscope.

Note: LXI, GPIB, LSIB, and USBTMC remote control is available for some model instruments.
Follow the same procedure, selecting the appropriate connection type on the oscilloscope and
in QualiPHY.

Configure Oscilloscope for Remote Control

1. From the menu bar, choose Utilities

Utilities Setup...

2. Open the Remote tab and set Remote Control to TCP/IP.

3. Verify that the oscilloscope shows an IP address.

Add Connection to QualiPHY

1. On the host PC, download and run QualiPHYInstaller.exe.

2. Start QualiPHY and click the General Setup button.

3. On the Connection tab, click Scope Selector.

4. Click Add and choose the connection type. Enter the oscilloscope IP address from Step

3 above. Click OK.

5. When the oscilloscope is properly detected, it appears on the Scope Selector dialog.

Select the connection, and click OK.

QualiPHY is now ready to control the oscilloscope.

Select Connection

Multiple oscilloscopes may be accessible to a single remote host. In that case, go to General
Setup and use the Scope Selector at the start of the QualiPHY session to choose the correct
connection.

QualiPHY tests the oscilloscope connection when starting a test. The system warns you if there
is a connection problem.