Setup menu – TIF 3340 Professional Differential Thermometer and InfraRed Thermometer User Manual

Page 4

Advertising
background image

Set the Sample Rate of
Temperature Data Recording

The default sample rate is one

second, which is displayed on

the secondary display as 0001

(minute–second).

1. While in Setup Menu, press to select

which reading (minute—second) to adjust.

2. Press or until desired setting displays.

3. Press ENTER to store the new setting in

memory.

SETUP MENU

000 1

K

SETUP

Set the Time

The secondary display shows

year, month, and day. The

tertiary display shows hour and

minute. June 10, 2009, 1:31 PM

is displayed in the graphic.

2009 06 10

1331

K

SETUP

1. While in Setup Menu, press to select

which reading to adjust: year—month—day—

hour—minute.

2. Press or until desired setting displays.

3. Press ENTER to store the new setting in

memory.

8888

K

SETUP

Erase the Memory

The number of events stored is

shown on the secondary display.

This procedure outlines how to

clear the memory.

1. While in Setup Menu, press or until

CLR is displayed.

2. Press ENTER.

3. Press or until YES is displayed.

4. Press ENTER to clear the memory.

Back, Sheet 2

Option

Menu Item

Settings

Offset

T1, T2, IR, T3, T4 T1, T2, IR, T3, or T4 offset

Sample Rate rAt

Recording sample rate

Time Setting tiE

Show / set the time

Clear

CLR

Clear memory of events

Sleep Mode

SLP

ON or OFF

Using Offset to Adjust for Probe Errors

The offset option can adjust thermometer

readings to compensate for the errors of a

specific thermocouple and IR temperature. The

adjustment range allowed is: ± 9.0°F or ± 5.0°C.
1. While in Setup Menu, plug the thermocouple

into the input connector.

2. Place the thermocouple in a known, stable

temperature environment (such as an ice dry

well calibrator). Allow the readings to stabilize.

4. Press or until the primary reading

matches the calibration temperature.

3. Press ENTER to store new setting in memory.

Note: The primary display shows the temperature

plus the offset, and the secondary display shows

the offset. Individual offsets can be stored for T1,

T2, IR, T3, or T4.

CHANGING SETUP OPTIONS

1. Press for three (3) seconds to enter

Setup Menu. SETUP appears at the bottom of

the display.

2. Press or to scroll to the desired setup

option.

3. Press ENTER to change this setting.

4. Press or until desired setting displays.

5. Press ENTER to store new setting in memory.

6. Press SETUP for three (3) seconds to exit

Setup Menu.

NOTE: Setup is disabled in MAX MIN AVG mode.