REV 1 EZCT-2000B USER’S MANUAL

7

Table 2. Functional Descriptions of EZCT-2000B Controls and Indicators

Item

Number

Panel Markings

Functional Description

1

INSULATION

RESISTANCE

METER

Insulation resistance meter connectors.

2-6

X1, X2, X3, X4, X5

Current transformer excitation voltage connectors. Each set of connectors
contains a test voltage connector and sensing connector. The EZCT-2000B’s X
output terminals are rated to 2000 Vac working voltage. Any voltage above
2000 Vac will damage the input circuitry
.

7

H

Current transformer primary input test cable connector.

8

GROUND

Grounding stud.

9

None

AC receptacle.

10

100-120Vac, 12A

50-60Hz

Power switch with built-in circuit breaker.

11

None

Back-lit graphic LCD screen (128 x 64); viewable in bright sunlight and low light
levels.

12

None

Rugged, alpha-numeric membrane keypad

13

HIGH VOLTAGE

PRESENT

LED warning indicator that is illuminated when high voltage is present.

14

EMERGENCY

TURN OFF

“PUSH”

Emergency turn off switch.

15

USB MEM

USB Flash drive interface.

16

USB PC

USB PC interface.

17

RS-232C

RS-232C PC interface. Baud rate is set for 115KB, 8 data bits, 2 stop bits.

18

None

Built-in 4.5-inch wide thermal printer.

19

AC CURRENT

SOURCE

0 – 20A

AC current source connectors.