Auto Charge in Advisory Mode

When enabled and a shockable rhythm is detected after pressing the ANALYZE button, the
unit will automatically charge the defibrillator to the pre-configured or user selected energy
setting.

Enable Voice In Advisory Mode

Enables or disables voice prompts when operating in Advisory or Manual operating modes. If
enabled, the voice prompts will sound at the appropriate time when in Advisory mode or
Manual mode.

ECG Waveform to Card in Advisory Mode

Sets the unit to record the ECG waveform continuously to the PCMCIA card while in advisory
mode.

Audio Data Recording to Card in Advisory Mode

Sets the unit to record audio data continuously to the PCMCIA card while in advisory mode.

Note:

Audio recording is temporarily suspended during CPR monitoring.

Display Heart Rate in Semi-Auto Mode

Sets the unit to display the patient heart rate on the monitor while in semiautomatic mode.

Display Lead on Power-Up

Allows the user to select any one of the following Leads to be displayed on power-up: Paddles/
Pads, I, II, or III. When the unit powers up in semi-automatic mode, the option is ignored. The
unit will power up in Lead II, and switch leads automatically between leads II and MFE for
Auto Monitor and Auto Defib modes, respectively. When the unit enters Manual Mode (Manual
Mode override sequence is completed), the lead switches to the setting specified in the “Display
Lead on Power Up” option.

Frequency Response

Selects the ECG filter bandwidth to be used during ECG monitoring. The frequency of the
device can be set to either help reduce noise, help with the diagnostic response, or set as a
default response.

Enable 12-Lead in Semi-Auto Mode (AED with 12-Lead)

If set to “Yes,” 12-lead analyses, transmissions and print-outs are available in semi-automatic
(AED) Auto Monitor mode (ECG leads on the patient with MFE Pads disconnected), depending
on other configuration option settings. If set to “No,” 12-lead functions are available in Manual
Mode only.

QRS Volume on Power-Up

Turns the default QRS beeper volume to midrange or off following power up of the unit.

If set to “Off in Semi-Auto Mode,” the QRS beeper volume turns off in semi-auto mode. If the
unit is changed to manual mode, the beeper volume sets to midrange.

Pace Rate Setting on Power-Up

Sets the default pace rate of the M Series unit on power-up, between 30 ppm and 102 ppm
(increments of 4 ppm).