ZOLL M Series Defibrillator Rev H User Manual

Page 7

Advertising
background image

12-lead

9650-0215-01

12-lead - 5

If the M Series unit has been configured to automatically
transmit 12-lead ECG data records to a fax machine or Catalyst
MUSE system following completion of the acquisition phase,
the following data transmission screen will be displayed.

Pressing the Prev Phone# and Next Phone# softkeys will
scroll the highlight up and down through the preconfigured
phone destinations

Pressing the Dial Phone# softkey will dial the highlighted
phone destination and begin the transmission process.

Pressing the Manual Dial softkey will display the following
screen for entering the phone number:

Pressing the Prev Digit softkey moves the highlight left to the
previously entered digit.

Pressing the Inc Digit or Dec Digit softkey increments or
decrements the selected digit.

Pressing the Next Digit softkey moves the highlight right to the
next digit to be entered.
If a minimum of one (1) digit has been entered, pressing the
Dial Phone# or Return softkey will dial the entered phone
number and display the following screen:

Standard modem call progress status messages will follow as
the various phases of the transmission progress.
If you are transmitting to a fax machine, the following messages
display in sequence: “FAX PREPARING,” “FAX DIALING,”
“FAX SENDING,” “FAX DONE”.
If you are transmitting to a Catalyst MUSE system, the following
messages display in sequence: “MUSE DIALING,” “MUSE
SENDING,” “MUSE DONE.”

Pressing the Abort softkey will abort the transmission process.
Turning the Selector Switch to another mode of operation
(Defibrillation or Pacing) will also automatically abort the
transmission.
When the transmission is complete, an error has occurred or
the Abort softkey is pressed, the Retry screen will be
displayed:

Pressing Retry will display the Transmission screen, so that
transmission can be performed again. Pressing Return will
display the main 12-lead screen. If 30 seconds elapse and a

softkey is not pressed, the main 12-lead screen will be
displayed.

Fault Conditions

Transmission Error

In the event of a failure to transmit the ECG record, the
operator will be notified and the M Series will continue to retry
the transmission until the operator aborts the transmission or
the transmission is successful. See “Troubleshooting” for
possible transmission error messages and corrective actions.

Check Recorder

If an out of paper condition or a paper jam condition occurs
during printout of the 12-lead report, the operator will be notified
but data transmission will continue. The operator can retrieve
the 12-lead report by printing a Summary Report after the error
condition is cleared.

Settings Softkey

Pressing the Settings softkey from the 12-lead Monitor screen
allows the user to change ECG Filter, Lead Group settings or
Dial Type for 12-lead monitoring. ECG Filter and Lead Group
setting defaults can be programmed via the Configuration mode
of the M Series. (See the M Series Configuration Guide, P/N
9650-0201-01,
for more information.)

Once the settings screen is displayed, pressing the Settings
softkey will scroll the highlighted area among the different
available selections. Pressing Enter selects the setting that is
highlighted. The Return softkey allows the user to return to the
12-lead Monitor screen.

AED units in Semi-Automatic monitor mode do not display
“Filter” or “Lead Grp.” Only “Dial Type” displays.
Filter Setting

Selecting the “Filter” setting allows the user to select among the
following three filter / print format settings for 12-lead monitoring
ONLY :
0.05 – 150 4x3 channel printout (0.05 – 150Hz 4x3) : When
programmed to this setting, all 4x3 12-lead ECG reports will
print with a 0.05 to 150 Hz bandwidth.
0.05 – 150 Continuous printout (0.05 – 150Hz Cont.) : When
programmed to this setting, pressing and holding the Recorder
button will print a continuous ECG strip using full diagnostic
bandwidth until the Recorder button is released. (This is an
alternative to producing a 4x3 12-lead report in this fashion. A
4x3 12-lead can still be printed using the Acquire softkey.)
Note: Use a diagnostic bandwidth of 0.05 to 150 Hz for
diagnostic ECGs. ZOLL suggests using a 0.05 to 40 Hz setting
in environments prone to noise.
0.05 – 40 4x3 channel printout (0.05 – 40Hz 4x3) : When
programmed to this setting, all 4x3 12-Lead ECG reports will
print with a 0.05 to 40 Hz bandwidth. This setting can be used
to reduce excessive muscle or other artifact. This setting does