Test and troubleshooting, Calibration, 1pa100 – Pacific Laser Equipment 1PA-100 User Manual

Page 7

Advertising
background image

Amplifier 1ch100

User manual

1PA100

October 10, 2004

Release: 1.30

2002 by

Pacific Laser Equipment

Test and troubleshooting

High voltage output

Attention:

The HV output can reach +150V. This is a dangerous voltage.
Test in RC mode. At 5V analog input the indicator should read MAX.

Fuse

Check the main fuse inside the unit. When replacing it, use slow-blow 100mA.
The fuse is located inside the box and access to it is fairly difficult.

AC Power

Make sure that the AC power used is always 120V.

Calibration

Adjusted during manufacturing RC mode with the front panel potentiometer fully turned
counterclockwise and 0V analog input, the HV output is 0V.
At 5V analog input the HV output is 150V DC.

4.

Calibration

PLEquipment, 3941 S. Bristol St. Unit E, # 122, Santa Ana, CA. 92704,

800 984-3482, Fax: 509-355-5155, [email protected]

Page

7

of 9

Advertising