Specifications
Analog Inputs
Thermocouple Ranges
RTD Ranges
Analog inputs
8, solid state, differential
Optical isolation
120 Vac between inputs; 175 Vac from input to ground
A/D converter
Integrating voltage to frequency
DC voltage range
-10 to 60 mV. You can change it with scaling resistors to
any range from 0 to 25V
Resolution
0.02%, greater than 12 bits
Accuracy
0.05% at 25°C
Calibration
Automatic zero and full scale
Temperature coefficient
Less than 50 ppm/C, 0.005%/°C
Normal mode rejection
60 db at 60 Hz, full scale range maximum
Loop update time
2 times per second, every input
T/C burnout
Full upscale reading of input standard
DC milliamp inputs
4-20 mA, 10-50 mA, 0-50 mA, etc., with scaling resistors
Input Range
0-1V, 0-5V, 0-10V, 0-12V, up to 0-25 Vdc, with scaling
resistors
Infrared inputs
Power supply included, with scaling resistors
Source impedance
Measurements are within specification with up to 500
ohms source resistance
J
-350 to 1400°F -212
to
760°C
K
-450 to 2500°F
-268 to 1371°C
T
-450 to 750°F
-268 to 399°C
B
150 to 3200°F
66 to 1760°C
S
0 to 3200°F
-18 to 1760°C
R
0 to 3210°F
-18 to 1766°C
N
-450 to 2370°F
-268 to 1299°C
RTD1
-148.0 to +572.0°F
-100 to 300°C
0.1°
RTD2
-184.0 to +1544.0°F
-120 to 840°C
0.2°