<4. WIRING>

4-2

IM 01E25D11-01EN

4.1.2 Cables

(1) Dedicated Signal Cable (AXFC)

Outer jacket

10.5 (0.413")

Tape

Shield (C)

Insulation

Insulation

Shields (SA and SB)

Conductors (A and B)

F0401.ai

Figure 4.1.1 Dedicated Signal Cable AXFC

The flow signal is transmitted via this dedicated

cable. The cable is constructed with double

shielding over the two conductors, and heat-

resistant vinyl is used for the outer jacket material.

Finished diameter:

10.5 mm (0.413”)

Maximum length:

Combination with the AXFA11 converter:

200 m (660 ft)

Maximum temperature: 80°C (176°F)

IMPORTANT

If the cable is longer than required, cut off

any extra length rather than coiling it up, and

terminate the conductors as shown in Figure

4.1.2. Avoid using junction terminal boards to

extend the cable length, as this will interrupt the

shielding.

20 (0.8)

ø10.5 (0.4)

AXFC

L (Specified Dimensions)

150

±5

8 (0.3) max.

150

±5

70 (2.76)

60 (2.36)

90 (3.54)

(5.9)

(5.9)

55 (2.17)

90 (3.54)

50 (1.97)

25 (0.98)

SA

SB

A

C

B

A

C

B

8(0.3) max.

White

Black

Red

White

Black

Red

F0402.ai

Unit : mm

(approx. inch)

On the

converter

side

On the

flowtube

side

Figure 4.1.2 Treatment of Dedicated Signal Cables

CAUTION

As crimp terminals A, B, SA, SB and C have

their own electrical potentials, securely

insulate them so as not to come in contact

with one another.

To prevent a shield from coming in contact

with another shield or the case, cover each

shield with a vinyl tube or wrap it in vinyl tape.

NOTE

Conductors A and B carry the signal from the

electrodes, and C is at the potential of the liquid

(signal common). Shields SA and SB are kept at

the same potentials as the individual electrodes

(these are actively driven shields.) This is done

to reduce the effect of the distributed capacitance

of the cable at long cable length. Note that, since

the signals from the individual electrodes are

impedance converted inside the converter, errors

will result if they come in contact with any other

component. Great care must be taken in the

cable finish treatment.

(2) Excitation Cable
JIS C 3401 control cable equivalent

JIS C 3312 power cable equivalent

14 AWG Belden 8720 equivalent

Outer Diameter: 6.5 to 12 mm (0.26 to 0.47 in.)

10.5 or 11.5 mm (0.41 to 0.45 in.) for optional

code EG, EU and EW.

Nominal Cross Section:

Single wire: 0.5 to 2.5 mm

2

Stranded wire: 0.5 to 1.5 mm

2