<3. INSTALLATION>

3-7

IM 01E30D01-01EN

Table 3.3.3 Wafer Type Tightening Torque Values for Metal Piping and Permeable Fluids

Tightening torque values for PFA lining type (N-m / {kgf-cm} / [in-lbf])

Gasket types

within flowtube

No gasket (standard)

Gasket types

for user’s

flange

PTFE-sheathed non-asbestos gasket (optional codes BCF and BSF), or the equivalent in hardness

Flange ratings

Size

mm (inch)

JIS 10K, ANSI Class 150, and DIN PN10

JIS 20K, ANSI Class 300, and DIN PN16

DIN PN40

25 (1.0)

34.9 to 40.1 / {355.9 to 408.9} / [308.9 to 354.9] 35.2 to 40.1 / {358.9 to 408.9} / [311.5 to 354.9] 32.3 to 37.1 / {329.4 to 378.3} / [285.9 to 328.4]

40 (1.5)

53.5 to 61.5 / {545.5 to 627.1} / [473.5 to 544.3] 54.2 to 61.5 / {552.7 to 627.1} / [479.7 to 544.3] 56.4 to 61.5 / {575.1 to 627.1} / [499.2 to 544.3]

3.3.2 Nominal Diameter 50 mm (2.0 in.) to

200 mm (8.0 in.), Wafer Type

IMPORTANT

Use bolts and nuts in compliance with the flange

ratings. When stud-type through-bolts are used,

be sure the outside diameter of the shank is

smaller than that of the thread ridge. Be sure to

choose a gasket with inner and outer diameters

that does not protrude inside the piping (refer to

Table 3.3.9). If the inner diameter of the gasket is

too large, or outer diameter of the gasket is too

small, fluid leakage may result.

(1) Mounting Direction
Mount the flowmeter so that the flow direction of the

fluid to be measured is in line with the direction of

the arrow mark on the flowmeter.

IMPORTANT

If it is impossible to match the direction of

the arrow mark, the direction of the electrical

connection can be changed. Refer to Section

10.1 to do this properly.

In case the fluid being measured flows against

the arrow direction, refer to this parameter

J20:

Flow Direction in this user’s manual.

(2) Mounting Centering Devices
To maintain concentricity of the flowmeter with the

pipes, install centering devices. From the process

piping side, pass two through-bolts through the

four centering devices (two for each bolt) and

the adjacent two holes (the lower two holes for

horizontal mounting) of both of the flanges (refer

to Figure 3.3.2). Use the appropriate centering

devices according to the nominal diameter and the

flange ratings. The centering devices are engraved

with an identifying character. Use the appropriate

ones which meet the required specifications by

referring to Tables 3.3.7.

(3) Positioning the Flowmeter
Position the flowmeter so that the Mini-flanges and

the centering devices come in close contact with

each other. Be careful to prevent the four centering

devices from coming into contact with the housing.

If stud-type through-bolts are used, position them in

such a way that the four centering devices come in

contact with the bolt threads (refer to Figure 3.3.2).

Pass the other through-bolts through from the

process piping side.