4 simulation parameter setup by fieldmate r1.03, Simulation parameter setup by fieldmate r1.03 -9 – Yokogawa AXFA14G/C User Manual

Page 52

Advertising
background image

IM 01E20F02-01E

6-9

6. IN-PROCESS OPERATION

6.3.4 Simulation Parameter Setup by

FieldMate R1.03

This section describes the procedure taken to set and
change the parameters in a use of simulation function
in each function block.

IMPORTANT

• Before running the program, log-in to Windows

as an Administrator or as a user with adminis-
trative authority.

• When comm. DTM is connected to the PROFI-

BUS network with class 1 master running, set
the same busparameters of class 1 master.

It is mandatory to change the [Target Mode] from
[AUTO] to [O/S] before using this function in each
function block. Also the hardware slide switch (No 2)
in the AXF amplifier must be set “ON” beforehand
when this function is used.

The hardware slide switch and the [Target Mode] in
each function block must be returned “OFF” / “AUTO”
when the simulation operation was finished.

IMPORTANT

Do not turn power OFF immediately after param-
eter setting. When parameters are saved to the
EEPROM, the redundant processing is executed
for an improvement of reliability. Should the
power be tumed OFF within 60 seconds after
setting of parameters, changed parameters are
not saved and may return to their original values.

(1) Simulate Measured Value Setup in the

Transducer Block

Entry simulation, its type, its values at [Simulate
Measured Value] in the [Transducer Block] menu.

Figure 6.3.4

Simulate Measured Value Setup

(2) Simulation Setup in the AI Block

Entry the simulation value, simulation quality, status
and simulation in the [Analog Input] menu.

Figure 6.3.5

Simulation Setup

(3) Simulation Setup in the TOT Block

Select adequate [Simulation Value], [Simulation
Quality], [Simulation] and [Status] in each [Totalizer]
menu.

Figure 6.3.6

Simulation Setup