2 fm (usa + canada), 1 technical data, 2 fm (usa + canada) -9 – Yokogawa RotaMASS 3-Series User Manual

Page 201: 1 technical data -9

Advertising
background image

9. EXplOSION prOTECTEd TYpE INSTrUMENTS

9-9

All Rights Reserved. Copyright © 2003, Rota Yokogawa

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

9.2 FM (USA + Canada)

Applicable standards:
FM3600, FM3610, FM3615, FM3810,
ANSI/NEMA 250, IEC 60529,
ANSI/ISA 12.00.01, ANSI/ISA 12.22.01,
CSA-C22.2 No.157, CSA-C22.2 No.25,
CSA-C22.2 No.30, CSA-C22.2 No.0.5,
CSA-C22.2 No.142, CSA-C22.2 No.0.4,
CSA-C22.2 No.94

9.2.1 Technical data

remote detector rCCS30 ... 39/Xr

(option /FS1):

- Intrinsically safe
- AEx ia IIC, Class 1, Zone 0
- IS Class I, Division 1, Groups A,B,C,D T6
- DIP Class II / III, Division 1, Groups E,F,G
- IP67 / NEMA 4X
remote converter rCCF31 (option /FF1):
- Housing explosion proof
- Provides intrinsically safe detector circuits
- AEx [ia] IIC, Class I, Zone 1, T6
- Class I, Division 1, Groups A,B,C,D
- Class I, Division 1, Groups C,D with
option /HP
- Class II / III, Division 1, Groups E,F,G
- AIS Class I / II / III, Division 1, Groups

A,B,C,D, E,F,G
- AIS Class I / II / III, Division 1, Groups

C,D,E,F,G with /HP
remote converter rCCr31 (option /FS1):
- Intrinsic safe associated apparatus
- Provides intrinsically safe detector circuits
- [AEx ia] IIC, Class I, Zone 1
- AIS Class I, Division 1, Groups A,B,C,D
- AIS Class I, Division 1, Groups C,D with

option /HP
Integral type rCCT34 .. 39/Xr (option /FF1):
- Housing explosion proof
- AEx d [ia] IIC, Class I, Zone 1, T6
- Class I, Division 1, Groups A,B,C,D
- Class I, Division 1, Groups C,D with
option /HP
- Class II / III, Division 1, Groups E,F,G
- IP67 / NEMA 4X
Temperature classification :
The remote converter RCCF31 has a T6 temperature
class rating for operation at ambient temperature
up to +50°C / +122°F.

Process temperature limits / Heat carrier fluid
temperature limits :
- Standard : -50°C to 150°C / -58°F to 302°F
- with option /MT: -50°C to 220°C / -58°F to 428°F
- with option /HT: 0°C to 350°C / -32°F to 662°F

Electrical data remote converter rCCF31,
rCCr31 and converter of integral type rCCT3 :
- Driving circuit : terminals D+ / D-

Uo = 14.5 V; Io = 47 mA; Po = 0.171 W

Lo = 15 mH; Co = 0.65 µF

- Driving circuit : terminals D+ / D- with option /HP

Uo = 11.7 V; Io = 124 mA; Po = 0.363 W

Lo = 8 mH; Co = 10.3 µF

- Sensor circuits: terminals S1+/ S1- or S2+ / S2-

Uo = 14.5 V; Io = 47 mA; Po = 0.363 W

Lo = 15 mH; Co = 0.65 µF

- Temperature sensor circuit : terminals

TP1,TP2, TP3

Uo = 13.3 V; Io =40 mA; Po = 0.133 W

Lo = 20 mH; Co =0.91 µF

Electrical data remote detector rCCS30 ... 33:
- Driving circuit : terminals D+ and D

Ui = 16 V; Ii = 53 mA; Pi = 0.212 W

Li = 4.2 mH; Ci = negligible small

- Driving circuit : terminals D+ / D- with option /HP

Ui = 16 V; Ii = 153 mA; Pi = 0.612 W

Li = 4.2 mH; Ci = negligible small

- Sensor circuits: terminals S1+ / S1- or S2+ /S2-

Ui = 16 V; Ii = 80 mA; Pi = 0.32 W

Li = 4.2 mH; Ci = negligible small

- Temperature sensor circuit : terminals TP1,
TP2, TP3

Ui = 16 V; Ii =50 mA; Pi = 0.2 W

Li = negligible small; Ci = negligible small

Electrical data remote detector rCCS34 ... 39/Xr :
- Driving circuit : terminals D+ and D

Ui = 16 V; Ii = 53 mA; Pi = 0.212 W

Li = 3.2 mH; Ci = negligible small

- Driving circuit : terminals D+ / D- with option /HP

Ui = 16 V; Ii = 153 mA; Pi = 0.612 W

Li = 3.2 mH; Ci = negligible small

- Sensor circuits: terminals S1+ / S1- or S2+ /S2-

Ui = 16 V; Ii = 80 mA; Pi = 0.32 W

Li = 2.1 mH; Ci = negligible small

- Temperature sensor circuit : terminals TP1,
TP2, TP3

Ui = 16 V; Ii =50 mA; Pi = 0.2 W

Li = negligible small; Ci = negligible small

Advertising