Explosion proof specifications, Caution – Yokogawa YFGW710 User Manual

Page 10

Advertising
background image

10

<Read Me First>

IM 01W01F01-11EN

Explosion Proof Specifications

CAUTION

This instrument has been tested and certified as being explosion proof. Please note that severe

restrictions apply to this instrument’s construction, installation, external wiring, maintenance and

repair. A failure to abide by these restrictions could make the instrument a hazard to operate.

n

Optional Specifications(For Explosion Protected Types)

Item

Description

Code

Factory Mutual

(FM)

FM nonincendive approval

Applicable standard: FM3600, FM3611, FM3810, ANSI/NEMA250

Nonincendive for Class I, Division 2, Groups A, B, C and D;

Class II, Division 2, Groups F and G;

Class III, Division 1 hazardous (Classified) locations.

Enclosure: NEMA 4X

Temp. Class: T4, Amb. Temp.: –40 to 60°C (–40 to 140°F)

FN17

ATEX

ATEX Type n declaration

Applicable standard: EN60079-0: 2009, EN60079-0: 2012,

EN60079-11: 2012, EN60079-15: 2010, EN60079-28: 2007

II 3G Ex nA [ic] [op is] IIC T4 Gc (with 100BASE-FX)

II 3G Ex nA [ic] IIC T4 Gc (without 100BASE-FX)

Enclosure: IP66

Amb. Temp.: –40 to 60°C (–40 to 140°F)

KN27

Canadian Standards

Association

(CSA)

CSA nonincendive approval

Certificate: 2342292

[For CSA C22.2]

Applicable standard: C22.2 No.0, C22.2 No.0.4, C22.2 No.25,

C22.2 No.94, C22.2 No.213, C22.2 No.61010-1-04

Nonincendive for Class I, Division 2, Groups A, B, C and D;

Class II, Division 2, Groups F and G;

Class III, Division 1

Enclosure: Type 4X

[For CSA E60079]

Applicable standard: CAN/CSA E60079-0, CAN/CSA E60079-15,

IEC60529

Ex nA nL IIC T4

Enclosure: IP66

Amb. Temp.: –40 to 60°C (–40 to 140°F)

CN17

IECEx Scheme

IECEx Type n approval

Certificate: IECEx DEK 13.0047

Applicable standard: IEC60079-0:2011, IEC60079-15:2010,

IEC 60079-11:2011

Ex nA [ic] IIC T4 Gc

Enclosure: IP66

Amb. Temp.: –40 to 60°C (–40 to 140°F)

SN27