Appendix b – Yokogawa BT200 User Manual

Page 51

Advertising
background image

A-4

<Toc> <Ind>

<APPENDIX>

IM 01C00A11-01E

APPENDIX B

Installation of CSA Intrinsically Safe Type

BT200 BRAIN TERMINAL (Model BT200-N00/CS1)
is applicable for use in hazardous locations;

Applicable Standard:

For CSA C22.2 Series;

C22.2 No 0, C22.2 No 142, C22.2 No 157, C22.2
No 1010.1

Certificate: LR 81741C-33

Intrinsically Safe for Class I, Division 1, Groups A,
B, C & D Hazardous Locations.

Temperature Class: T4

Ambient Temperature: -15 to 55

°C

Electrical Parameters

BT200 Intrinsically Safe Apparatus Parameters
The Maximum Input Voltage

Vmax(in)=30V

The Maximum Input Current

Imax(in)=165mA

The Maximum Input Power

Pmax(in)=0.9W

The Maximum Internal Capacitance

Ci=0

The Maximum Internal Inductance

Li=730

µH

The Maximum Output Voltage

Vmax(out)=2V

The Maximum Output Current

Imax(out)=22mA

The Maximum Output Power

Pmax(out)=11mW

The Maximum Allowed Capacitance

Ca=3000

µF

The Maximum Allowed Inductance

La=30mH

Hazardous Location

Non-hazardous Location

+

+

+

Intrinsically Safe
Apparatus

Safety Barrier

F0B01.EPS

General Purpose
Equipment

+

+

BT200

Advertising