(3) Using the values in Table 6.2, check the MLX readings by setting the current source to

each of the table values. Check all points in the table and verify unit is within specification.

Table 6.2 Percent vs. Current Values

Value

Current

0%

4.0mA

25%

8.0mA

50%

12mA

75%

16mA

100%

20mA


(4) If a re-calibration is required, use the following method:

(a) Touch the “Menu” button to enter the menu system.

The alphanumeric display reads “Setup”.

(b) Touch

to begin the setup routine.

The alphanumeric display reads “Calibrate”.

(c) Touch

to begin the calibrate routine. Follow the instructions given on the

display. For each calibration point, after adjusting the current standard to the
value indicated, press the “Menu” button.




6.4 Rotating Display Direction

The MLX display is designed so that it can be rotated in increments of 90 degrees. When there is
a need to change the orientation of the display, use the following procedure:

(1) Remove power from the MLX.
(2) Remove the glass cover from the display side.
(3) Remove the two anchor nuts from the MLX module assembly.
(4) Remove and rotate the display assembly to the desired orientation.
(5) Re-install and tighten the two nuts on the module assembly.
(6) Replace the glass cover.

6.5 Cleaning

Cleaning should be restricted to wiping with a damp cloth or approved anti-static cleaner to avoid
the danger of ignition due to electrostatic charges.