<3. Functions>

3-3

IM 11B06F01-01E

Coil (00001 to 09999)

Data

Address

Description

DO ON com-

mand

00101 to 00116 A 1 is set when DO001 to DO016 is ON. The coil is back to 0 auto-

matically after setting a 1.

DO OFF com-

mand

00201 to 00216 A 1 is set when DO001 to DO016 is OFF. The coil is back to 0 auto-

matically after setting a 1.

Input relay (10001 to 19999)

Data

Address

Description

ASIU normal

10001

Neither Power Failure, I/O module error nor IOP status happens. All

of the address from 12001 to 12199 has 0 values.

ASIU error

10002

Either Power Failure, I/O module error or IOP status happens. One of

the addresses from 12001 to 12199 has 1 value.

DI (Digital

Input) read

10301 to 10316 DI01 to DI16

DO (Digital

Output) read

10401 to 10416 DO16 to DO16

Power Failure

12001

12002

A 1 is displayed when Main Power is failed.

A 1 is displayed when 24 V DC Power is failed.

I/O module

error

12011 to 12014 A 1 is displayed when the module is failed. The address is followed

the location of the I/O module. From left to right, 11, 12, 13 and 14 is

applied.

AI IOP status

12101 to 12116 It corresponds to AI01 to AI16. A 1 is displayed when it is IOP status.

Input register (30001 to 39999)

Data

Address

Description

AI read (Real

data)

30401 to 30432 The average value during 1 second with 10 scans times. However

when it is IOP status, IOP value, which is 0 or 1, is displayed. When

real number format is selected, the data is displayed with 2 words

address.

The data range is 0.0 to 1.0

AI read (9999)

30501 to 30516 The average value during 1 second with 10 scans times. However

when it is IOP status, IOP value, which is 0 or 9999, is displayed. It

means data normalized by 9999 (Scaling Coefficient).

The data range is 0 to 9999

AI read (65535) 30601 to 30616 The average value during 1 second with 10 scans times. However

when it is IOP status, IOP value, which is 0 or 65535, is displayed. It

means data normalized by 65535 (Scaling Coefficient).
The data range is 0 to 65535

Holding register (40001 to 49999)

Data

Address

Description

Not used

l

Access to the address not listed or mapped

Access to the address that is not shown on the table or not mapped does not generate an error.
However, the value of data is uncertain.
DCS can write on coil or holding register of ASGW that is not defined, however ASGW does not