<1. Wiring and Installation>

3

IM 12A01A02-12E

4th Edition : Sep. 3, 2013-00

1.3

Wiring the sensor

The FLXA21 can be used with a wide range of commercially available sensor types, both from

Yokogawa and other manufacturers. For more detailed information, refer to the respective

instruction manual of the sensor. The sensor systems from Yokogawa fall into two categories; the

ones that use a fixed cable and the ones with separate cables.
To connect sensors with fixed cables, simply match the terminal numbers in the instrument

with the identification numbers on the cable ends. The separate sensors and cables are not

numbered, but instead use a color-coding system.

1.3.1

pH Measurement

Conventional pH sensors are connected to the module as follows:

11 Temperature
12 Temperature
13 Reference
14 Solution ground
15 Glass (measure)
16 Shield
17 Shield

FLXA21

REF

TC

PH

Liquid

Earth

In addition to the wiring of the sensor, insure that a jumper for low-impedance sensor inputs is

installed. The jumpers can be found on the plastic sensor module cover and can be stored in the

lower level module wiring cover.

• pH Glass Electrode is a high impedance sensor input
• Special electrodes using 2 glass sensor (example: Pfaudler, SC24V) do not need jumpers.

Terminals 15-16 are identified as input 1 (High Impedance) and terminals 13-17 are defined as

input 2 (Low Impedance). For conventional pH sensors, the jumper is placed as illustrated:

16

15

19

17

13

18

14

12

11

PH

Input 1

Input 2

Glass sensor on Input 1.

Reference sensor on Input 2.

Jumper

holder

Advertising