Only for peek – Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity User Manual

Page 45

Advertising
background image

37

IM 12D06D05-01E

6

CALIBRATION

Cell constant: The nominal cell constant of

the sensor is 1.88 cm

-1

for the PEEK sensor

types and 3.00 cm

-1

for the PFA sensor. The

calibrated values are indicated on the cable

markers and the actual installation can change

this factor. If there is less than 25 mm spacing

between sensor and holder, in-situ calibration is

necessary to meet the specified accuracies.

D

non conductive

piping

conductive

piping

D i n m i l l i m e t e r s

0 10 20 30 40 50

0.90

Correction factor (x nominal C.C.)

0.95

1.30

1.25

1.20

1.00

1,05

1.10

1.15

Only for PEEK