Yokogawa DO402 Dissolved Oxygen Analyzer User Manual

Page 59

Advertising
background image

IM 12J05D02-01E

Parameter setting 5-23

Code Display

Function

Function detail

X

Y

Z

Default values

Contacts Settings
44

*D.TIME

Dead Time

Delay after setpoint passed 0- 2.0 s

0.2 s

*P.HYST

Hysteresis

Process value hysteresis

0-50 mg/l or 0-300 %

0.1 mg/l

*T.HYST

Temp hyst

Temperature hysteresis

0- 5 °C (0- 12 °F)

0.5 °C

45

*RANGE

Prop. control

Proportional range 0- 50 mg/l

5.0 mg/l

*PER.

Duty cycle

Period/ Duty cycle : 5- 100 s

10.0 s

*FREQ.

Frequency

Frequency control 50 - 120 per min.

70 p.m.

46

*tI.CNT

Integral time

Int. time for contacts 10 - 1999 s

100 s

47

*EXPIR

Time out alarm

Disable time out alarm

0

0

Enable time out alarm

1

*tE. mn

time out alarm after 0.5 - 199.9 min

15 min

48-49

Not used