Setting the Input Range

Configure the FX so that it measures temperature on measurement channel 1 and flow
rate on measurement channel 2.

Setting the Temperature Measurement Channel, the Input Range, and the Tag

Type T

thermocouple

Channel 1

0.0 to 200.0°C

FX

Setup Item

Description

Number in the Figure

Channel

Use channel 1.

1

Tag

TI-001

2

Sensor

Type T thermocouple

3

Input range

0.0 to 200.0°C

4

(1) Input Range

Press

MENU (switch to the setting mode).

Select the

Menu tab > Meas channel > Range, Alarm.

1

4

3

Setting Procedure

1.

Display the operation mode screen.

2.

Press

MENU once to display the setting menu.

3.

Press the

down arrow key twice to select Meas channel.

4.

Press the

right arrow key once.

Select the Range, Alarm.

Setting the Functions and Operations